5.1 External reference
60
Acqiris SA230E User's Manual
5.1 External reference
For applications for which the user wants to replace the internal clock of the acquisition module and
drives the ADC with an external source, an external reference signal can be used. The reference
signal can be entered into the ADC module by the dedicated
REF IN
connectors.
External reference (REF IN)
For applications that require greater timing precision and long-term stability than is obtainable from the
internal clock, a 10 MHz or a 100 MHz reference signal can be used.
The external reference is nominally at 10 MHz or 100 MHz. However, frequencies in a range will be
accepted. If your input is not at exactly the specified value, you must remember to compensate for the
difference in your application since the ADC module and the driver have no way to know about such
deviations.
Parameter
Value
Tolerance
Nominal Frequency
100 MHz or 10
MHz
±1 kHz
Signal level
-3 dBm to +3 dBm
Impedance
50 Ω
Coupling
AC
Minimum ampliture
440 mVpp (sinus)
Maximum power
2 mW
Maximum voltage
900 mVpp (sinus)
Table 5.1
- External reference specifications.
If synchronization between several ADC modules is required, the reference signal should be applied to
all of them.
5.2 Trigger modes and time-stamps
Trigger modes
As listed previously the trigger source can be:
the signal applied to an input channel (internal triggering)
an external signal applied to the TRG IN front panel input connector (external triggering)
a software trigger (See
How to generate a software trigger? (page 81)
a self-trigger (See
for this specific mode)