4.1 Baseline stabilization and digital offset
Acqiris SA230E User's Manual
51
4.1 Baseline stabilization and digital offset
This functionality is recommended for time-domain applications needing very stable baseline, e.g.
sequences of pulse.
The baseline fluctuation comes from low frequency noise (e.g. 1/f noise in ADC) or from the
measurement side-effects.
The baseline stabilization can be enabled in both digitizer mode or in real-time averaging mode.
In addition, a digital offset can be applied after the baseline stabilization.
Figure 4.1 -
Continuous baseline stabilization example.
Principle
The baseline correction is not enabled by default. When activated, the baseline stabilization algorithm
consists in the main following steps:
1.
Baseline estimation
: The estimated baseline is iteratively adapted to the input data.
The baseline is estimated from sampling of the incoming input data, except when a pulse is detected
(pulse above or below a configurable threshold).
2.
Baseline correction
:
The estimated baseline is subtracted from the (unfiltered) input data.
3.
(optional) Digital offset
:
A configurable offset can be applied.
Baseline correction calculation period.
There are two modes for baseline correction calculation (see Figure below).