background image

6

This is 231.5 mm P2S - the white “5” being in 

line.  

 

BTW: This is the perfect mounting distance for 

the FR-64s.

This is 295,00 mm P2S - suitable for a FR-66s.

Side-kicks: 
Reading the  

Vernier scale

 

The SMARTractor features a Vernier scale to  

allow precise read-out of the P2S with an  

accuracy of 5/100 mm.

This Vernier scale runs right to left - as does the 

metric scale with increasing distance of P2S.

If you are unfamiliar with a Vernier scale, please 

get accustomed to its operation.

The white numbers 1- 9 over the black back-

ground are 1/10 of a mm each. The white lines 

between numbers are 0.5/10 mm each - or  

5/100 mm each.

Whenever any of the white lines is absolutely  

in-line with a black line on the aluminum scale, 

then it denominates the 5/100 value which has to 

be added to the full mm value given between the  

2 white “leaves” centered by the full vertical 

white line on the left of the scale just before the 

half-circle end.

To illustrate the read-out procedure, here are a 

few examples of mounting distances for common 

tonearms.

One will become accustom quite quickly to this 

way of reading the scale.

The Vernier scale is an international standard in 

industrial precision analog measuring. 

The magnifier

Using the glass magnifier greatly increases the 

visibility of the stylus when trying to align on 

the single spot. It further also allows with most 

modern cartridges to get a fairly good view of the 

cantilever.

Place the magnifier in the cut-out frame on the 

clear acrylic positioning arm and pull it all the 

way forward so that it seats squarely against the 

clear acrylic. 

It is automatically in a position to optimize focus 

when viewing through the glass lens on the  

alignment spot for the selected tangential curve.

The cutout for the glass magnifier in the  

acrylic positioning arm allows for adjusting the 

side-to-side position of the glass magnifier to  

accommodate all different positions of the 5  

tangential curves on the SMARTractor.

.

Содержание SMARTractor

Страница 1: ...Universal high precision triangular positioning alignment instrument for optimized analog phono tonearm alignment true one point cartridge set up owner s manual SMARTractor acoustical systems ...

Страница 2: ...rior to shipment Please do get yourself familiar to the handling and use of the SMARTractor While it is a mechanical positioning instrument it needs a minimum of attention to detail by the user to obtain the maximum precise results it ensures if handled correctly A brief introduction Most important Please study the step by step illustrated instruction to learn how to use and set up the SMARTractor...

Страница 3: ...SMARTractor an additional 20 mm of height for the positioning arm in those instance where required for illustration see page 8 The SMARTractor offers a total of 5 different tangential curves with which to align a given cartridge and tonearm combination Selecting the tangential curve You have the option to align the cartridge using any of these five tangential curves UNI DIN 1 Loefgren B DIN 2 Loef...

Страница 4: ...very hard time getting the white 20 mm POM ring off your spindle again The 12 mm and or 6 mm spindle adapter insert allows the precise centering of the SMARTractor on a wide range of spindle diameters Remember Please be careful and do not use any excessive force Pushing down realy hard creates a VERY tight lock which you will have big problems loosening up again Measuring P2S 20 mm spindle ring an...

Страница 5: ...ing where the exact pivot point is located This is ideal for precise positioning of the locator pin on the pivot point If there is no indication of the tonearm pivot point attempt to locate pin exactly over the tonearm pivot center If unsure of pivot center use the reticule to visually determine tonearm pivot point by finding symmetrical outer lines on the tonearm body Step 6 When the pivot point ...

Страница 6: ...e left of the scale just before the half circle end To illustrate the read out procedure here are a few examples of mounting distances for common tonearms One will become accustom quite quickly to this way of reading the scale The Vernier scale is an international standard in industrial precision analog measuring The magnifier Using the glass magnifier greatly increases the visibility of the stylu...

Страница 7: ...mounting screws after final position is obtained Step 9 When the stylus is precisely sitting on the single spot use the helping lines to ensure that the cantilever is perpendicular to the SMARTractor surface when viewed from the front Adjust the cartridge accordingly until the cantilever is visually in this position If you have a Uni Pivot design tonearm it is critical that the table is level befo...

Страница 8: ... Hofstetten Germany www acoustical systems com info acoustical systems com Step 10 Make sure that the stylus is precisely located on the spot with the helping lines straight with the cantilever making sure that the locator pin is still exactly positioned over the pivot When you are sure that the stylus is precisely located on the spot with the helping lines in line with the cantilever the tangenti...

Отзывы: