7
On Call
®
Sure
Blood Glucose Test Strips
The
On Call
®
Sure
Blood Glucose
Test Strips are thin strips with a chemical
reagent which work with the
On Call
®
Sure Sync
Blood Glucose Meter to measure
the glucose concentration in whole blood. After the strip is inserted into the meter,
blood is applied to the sample tip of the test strip. The blood is then automatically
absorbed into the reaction cell where the reaction takes place. A transient
electrical current is formed during the reaction and the blood glucose
concentration is calculated based on the electrical current detected by the meter.
The result is shown on the meter display. The meter is calibrated to display plasma
equivalent results.
IMPORTANT:
Apply the sample only to the sample tip of the test strip. Do not
apply blood or control solution to the top of the test strip as this may result in an
inaccurate reading.
Hold the blood drop to the sample tip of the
test strip until the check window is
completely full and until the meter begins
to count down. If you applied blood but do
not see the starting of the count down, you
may reapply a second drop of blood within
3 seconds. If the check window does not fill
and the meter starts to count down, then
do not add more blood to the test strip. If
you do then you may get an E-5 message
or an inaccurate test result. In this case if
the meter begins to count down and the
check window does not fill, discard the strip
and begin the test again with a new test
strip.
Sample Tip
Apply blood or control solution here.
Check Window
Check to confirm that a sufficient
sample size has been applied.
Contact Bars
Insert this end of the test strip
into the meter until it stops.
Incorrect
Correct
Correct
Incorrect