If the display stays at 

SEt   100

 after you press SET more than 

once, then the parts are too light 

7. If the sample met the proper criterion, then you will see a count on 

the display.  S

ee the chart below to determine the Minimum Piece 

and Sample Weights

8. Remove the sample from the platter 
9. If necessary, follow directions for entering a tare 
10. Place a sample of quantity to be counted 
11. Once the STABLE light is on you may read the pcs 
12. Remove sample from the platter and repeat steps 10 thru 12 as 

many times as necessary or proceed to step 13 

13. If necessary, follow directions for removing a tare

 

 

Parts Counting Specifications 

SI-1 

 

150 lb (60 kg) 

300 lb (150 kg) 

Minimum Piece Weight 

0.05 lb (0.02 kg) 

0.1 lb (0.05 kg) 

Minimum Sample Weight 

5 lb (2 kg) 

10 lb (5 kg) 

Maximum Quantity 

3,000 pcs 

 

Limit Mode 

(lb only)

 

1. Press the MODE key until you reach Limit mode (

Limit

2. Make sure that the scale is at ZERO and STABLE 
3. If necessary, follow directions for entering a tare 
4. Press the SET key 
5. The display will read 

Lo

 and then 

0.00

 or 

0.0

 

6. Place the Low Limit load on the platter and press the SET key 
7. The scale will display 

SEt

 momentarily and then display the 

weight of the Low Limit load 

8. Remove the Low Limit load from the platter 
9. Place the High Limit load on the platter and press the SET key 
10. The scale will display 

SEt

 momentarily and then display the 

weight of the High Limit load 

11. Remove the High Limit load from the platter 
12. You can now use the scale in Limit mode; the display will flash 

and scale will beep continuous when the weight meets the 
following criteria: 

0.00 < WEIGHT < Low Limit 
High Limit < WEIGHT < Max Capacity

 

13. If necessary, follow directions for removing a tare 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание SI-1

Страница 1: ...A A A ACOM COM COM COM OW N E R S M AN U AL MODEL SI 1 BENCH SCALE VER 1 00 ...

Страница 2: ... companies This company makes no representations or warranties either expressed or implied with respect to the contents hereof and specifically disclaims any warranties of merchantability or fitness for any particular purpose Any software described in this manual is sold or licensed as is Should the programs prove defective following their purchase the buyer and not this company its distributors o...

Страница 3: ... adjustable Units Modes pounds kilograms counting mode limit mode VFD Display 6 Digits 999999 max Platter Size Width 400 mm 15 7 Depth 525 mm 20 5 Width 400 mm 15 7 Depth 675 mm 26 6 Product Dimensions Height 850 mm 33 5 Input 120 Volts AC 60 Hz Power Requirements Output 9 Volts DC 300 mA Operating Temp 10 C 40 C 14 F 104 F Humidity RH 35 85 no condensation Power consumption Approximately 1 4 Watt...

Страница 4: ...ey no function Reserved for future use Display Zero lamp indicates that the scale is at net or gross zero Stable lamp indicates that the weight on the scale is stable Net lamp indicates that there is a tare set and that the displayed weight is the net weight When it is off it means that the displayed weight is the gross weight lb lamp indicates that the scale is displaying pounds kg lamp indicates...

Страница 5: ...Name Units QTY 3 1 IN CARTON BOX EA 1 3 2 STYROFOAM EA 2 3 3 BODY POLY BAG EA 1 3 4 MANUAL POLY BAG EA 1 3 5 AC ADAPTER EA 1 3 6 ADAPTER BOX EA 1 3 7 MANUAL EA 1 3 8 DISPLAY COLUMN EA 1 3 9 PLATFORM EA 1 3 10 INDICATOR EA 1 3 11 INDICATOR POLY BAG EA 1 3 15 SILICA DESICANT EA 2 ...

Страница 6: ... properly on top of the platform and that there is nothing on top of the platter Also make sure that the scale is plugged into the proper outlet Place the power switch into the ON position The display will run through a self test and then go to zero The scale will always remain in the last mode that it was left in The display may appear as follows it will read 0 0 if it is a 300 lb scale Changing ...

Страница 7: ...ZERO and STABLE Simple Weighing Using Tares 1 Make sure that the scale is at ZERO and STABLE 2 Follow directions for entering a tare 3 Place a load on the platter 4 After the STABLE indicator turns on you may read the weight 5 Remove the load 6 You may continue to use the same tare by returning to step 3 in this procedure OR continue to step 7 7 Follow steps for Removing Tares Parts Counting This ...

Страница 8: ...b 2 kg 10 lb 5 kg Maximum Quantity 3 000 pcs Limit Mode lb only 1 Press the MODE key until you reach Limit mode Limit 2 Make sure that the scale is at ZERO and STABLE 3 If necessary follow directions for entering a tare 4 Press the SET key 5 The display will read Lo and then 0 00 or 0 0 6 Place the Low Limit load on the platter and press the SET key 7 The scale will display SEt momentarily and the...

Страница 9: ...ing on or obstructing the platter Turn scale on with the platter properly placed on the scale EEP E2 PROM failure Contact Service AdC A D Converter failure Contact Service PArA Lost memory Possibly reprogram and recalibrate Contact Service Brand and product names are trademarks and or registered trademarks of their respective companies ...

Страница 10: ...Acom America Inc 404 Paulding Avenue Northvale NJ 07647 www acomamerica com Copyright 2001 All rights reserved 800 10006 020000 ...

Отзывы: