Installation
Page 33
Note:
When using multidrop ensure each instrument is given a unique
Serial ID
.
USB
Connect a standard type B USB cable (supplied with the instrument) to this port.
TCP/IP (optional)
Plug in an ethernet cable (this cable should be attached to a network).
Analog/Digital
This port is used to send and receive analog and digital signals. It is normally used to connect with a
gas calibrator or to activate alarm signals.
Each instrument contains eight digital inputs, eight digital outputs, three analog inputs and three
analog outputs.
Bluetooth
Connection is enabled using Ecotech’s Serinus Remote Android Application.
Use the Serinus Remote Android Application to access instrument and download data. It is available
for download directly from the Google Play Store. Search for “Ecotech Serinus Remote”.
2.3.4
Instrument Set-Up
1.
Open the lid and ensure the USB memory stick is installed (refer to Figure 6).
2.
Check the battery is turned on at the main controller PCA (refer to Figure 7).
3.
Turn on the instrument and allow the warm-up procedure to complete (refer to Section 3.1)
4.
Set the time and date (refer to Section 3.4.8).
5.
Set the digital filter to the desired setting (refer to Section 3.4.9).
6.
Set the internal data logging options (refer to Section 3.4.22).
7.
Set the analog/digital inputs and outputs settings (refer to Section 3.4.24).
8.
Perform a pressure sensor check (refer to Section 6.3.8).
9.
Perform a leak check (refer to Section 6.3.6).
10.
Leave the instrument to warm-up and stabilise for two-three hours.
11.
Follow the procedure for a multipoint precision check (refer to Section 5.7).
12.
The instrument is now ready for operation.
2.4
Transporting/Storage
Transporting the Serinus should be done with great care. It is recommended that the original
packaging material the Serinus was delivered in should be used when transporting or storing the
instrument.
Содержание Met One Instruments GAS-1060
Страница 10: ...Serinus 60 User Manual 3 0 Page 10 This page is intentionally blank...
Страница 18: ...Serinus 60 User Manual 3 0 Page 18 This page is intentionally blank...
Страница 35: ...Installation Page 35 Figure 7 Switching the Battery On Off...
Страница 36: ...Serinus 60 User Manual 3 0 Page 36 This page is intentionally blank...
Страница 102: ...Serinus 60 User Manual 3 0 Page 102 This page is intentionally blank...
Страница 122: ...Serinus 60 User Manual 3 0 Page 122 7 6 USB Memory Stick Failure Figure 64 USB Memory Stick Failure...
Страница 134: ...Serinus 60 User Manual 3 0 Page 134 9 5 Plumbing Schematic PN D020060...
Страница 135: ...Parts List Page 135 9 6 Block Wiring Schematic PN D020111...
Страница 136: ...Serinus 60 User Manual 3 0 Page 136 9 7 Measurement Cell Side Section PN H015000...
Страница 137: ...Parts List Page 137...
Страница 138: ...Serinus 60 User Manual 3 0 Page 138 9 8 Calibration Valve Manifold Assembly Exploded View PN H010013 01...
Страница 139: ...Parts List Page 139 9 9 Valve Assembly Exploded View PN H010042...
Страница 140: ...Serinus 60 User Manual 3 0 Page 140 9 10 High Pressure Valve Exploded View PN H050043...