1. General notes
This manual provides you with an overview of the necessary steps for commissioning and opera-
ting the computer system. Please read them even if you are familiar with the operation of compu-
ter systems.
We will use the term “unit” to designate the OR-PC
®
xx model manufactured by ACL GmbH. xx is
used to identify the display size between 09 and 55 inches.
2. Classification and intended use
The unit is designed for documenting data such as patient information inside and outside of
operating theaters, intensive-care wards and similar areas of application. It is supporting the local
software environment and it may be part of the IT network.
Following Section 6 of IEC 60601-1:2012 the unit is classified as below:
6.2 protective class I unit
6.3 see IP label on your unit, depending on your model IP20 or IP31
An optional IP65 upgrade is available on request for all models.
6.4 This unit is not intended for sterilization.
6.5 It is not intended for operating in an environment with enriched oxygen.
6.6 This unit is suited for permanent operation.
The unit is considered as professional ME-Equip according to IEC 60601-1-2:2014, paragraph 3.23.
It not is allowed to simultaneously touch the patient and pins of signal input/output components
when connecting the unit to other devices.
The unit can be used within the patient environment.
This unit is not suited to be used in direct contact with the patient.
The patient is not supposed to operate or maintain the unit.
It is not allowed to use this unit in private households.
The unit is not protected against explosions.
Only qualified expert personnel, which is familiar with the unit and its manual, may install, operate
and maintain the unit.
There is no information required for activities that should avoid damage to humans, animals or
environment.
Third-party equipment may be connected in consideration of the restrictions in
Section 3.1
.
3. Installation and commissioning
The unit was cleaned before leaving our production facilities. However, the units are neither sterile
nor disinfected on arrival due to transport conditions and packing. Advices for initial cleaning can
be found at
Section 5.1
.
Make sure that the unit and its accessories are complete according to the attached delivery note.
Keep the original transport box and packaging for the event that it might have to be transported
back (this does not apply to units with on-site service).
The integrated VESA-100 interface on the rear can be used to flexibly attach the unit to mounting
solutions and other devices. Various manufacturers offer a whole series of adapters or installation
solutions for this. We would be pleased to offer customized installation solutions on your request.
The unit may only be operated in an upright position (do not operate it in any direction at an
inclination of more than 30° from the norm). The only units that may be operated with a screen
rotated at 90° (portrait mode) are those specially designated as such (i.e., portrait models). It is not
allowed to operate it in impermissible positions (reversed and lying on its head, back or display)
because this may damage the unit.
Only use the screws supplied and always use all fastening screws (
orange
) for correct and save
assembly.