PRODUCT SPECIFICATIONS

Automation Components, Inc.

2305 Pleasant View Road  |  Middleton, WI 53562

Phone:

 1-888-967-5224  |  

Website:

 workaci.com

Page 4

Version: 1.0

I0000811

A/1.8K: 

1.8 KΩ nominal (Red/Yellow) 

A/3K:

 3 KΩ nominal (White/Brown)

A/AN (Type III):

 10 KΩ nominal (White/White)

A/AN-BC: 

5.238 KΩ nominal (White/Yellow)

A/CP (Type II):

 10 KΩ nominal (White/Green)

A/1.8K Series:

 +/- 0.5 °C @ 25 °C (77 °F)

 

 and (+/-1.0 °C) (+/-1.8 °F) 

A/100: 

100 Ω nominal 

+/- 0.06% Class A (Tolerance Formula: +/- °C = (0.15 °C + (0.002 * |t|))
 

where |t| is the absolute value of Temperature above or below 0 °C in °C)

@ -40 °C (-40 °F):

  +/- 0.23ºC (+/- 0.414ºF)

@ 0 °C (32 °F):

  +/- 0.15 °C (+/- 0.27 °F)

1 KΩ nominal (Orange/Yellow)

@ 21.1 °C (70 °F): 

+/- 1%

1 KΩ nominal (Red/Red)

@ -40 °C (-40 °F):

 +/- 1.52 °C (+/- 2.73 °F)

@ 0 °C (32 °F): 

+/- 0.4 °C (+/- 0.72 °F)

@ 21.1 °C (70 °F):

 +/- 0.17 °C (+/- 0.34 °F)

“-GD” Enclosure: 

Galvanized Steel, -40 to 121 °C (-40 to 250 °F), NEMA 1 (IP10)

“-PB” Enclosure:

 ABS Plastic, UL94-HB, -30 to 90 °C (-22 to 194 °F), Plenum Rated

“-BB” Enclosure:

 Aluminum, -40 to 121°C (-40 to 250 °F), Plenum Rated, NEMA 3R (IP 14)

“-4X” Enclosure:

 Polystyrene Plastic, UL94-V2, -40 to 70°C (-40 to 158°F), NEMA 4X (IP 66)

One
-40 to 85 °C (-40 to 185 °F)
0 to 90% RH, non-condensing
0.250” (6.35mm) | 0.260”
304 Stainless Steel | 304 Series Stainless Steel
22 AWG (0.65 mm)
½” NPS (National Pipe Straight) Female Thread | ½” NPT (National
 

Pipe Tapered) Male Thread

A/CSI: 

10 KΩ nominal (Green/Yellow)

A/10KS: 

10 KΩ nominal (White/Blue)

A/10K-E1:

 10 KΩ nominal (Gray/Orange)

A/20K:

 20 KΩ nominal (Brown/Blue)

A/100KS: 

100 KΩ nominal (Black/Yellow)

A/10K-E1 Series:

 +/- 0.3 °C (+/- 0.54 °F)

All Else:

 +/- 0.2 °C (+/- 0.36 °F) 

A/1K:

 1 KΩ nominal

@ 200 °C (392 °F):  

+/- 0.55 °C  (+/- 1.00 °F)

@ 54.4 °C (130 °F):

 +/- 0.56 °C (+/- 1. 00°F)

@ 121 °C (250 °F): 

+/- 1.25 °C (2.25 °F)

SENSOR NON-SPECIFIC INFORMATION

Number Sensing Points:
Storage Temperature Range:
Operating Humidity Range:
Probe Diameter | Thermowell Bore Diameter:
Probe Material | Thermowell Material:
Wire Size
Thermowell Instrument | Process Thread Size:

Enclosure Specifications

(Temperature, Material, 
Flammability, NEMA/IP Ratings)

THERMISTOR

Sensor Output @ 25 °C (77 °F):

(Lead Wire Colors)

*Does not include CL2P

 

Accuracy @ 0-70 °C (32 - 158 °F):

PLATINUM

Sensor Output @ 0 °C (32 °F):
Accuracy:

BALCO 

Sensor Output @ 21.1 °C (70 °F):

(Lead Wire Colors)

Accuracy:

NICKEL

Sensor Output @ 21.1 °C (70 °F):

(Lead Wire Colors)

Accuracy:

Отзывы: