Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 52 of 60
10.
Appendix
10.1.
Error Messages
If an error is detected, the error message, followed by a short description,
will be displayed. In this case, the normal operation of the device will not
respond to gas and the current loop output will be the ERROR value.
Display
Cause
Troubleshooting
PID lamp
The PID lamp does not work.
Check that the PID lamp
connector is well connected.
PID sensor
PID sensor does not work.
Check that the cable between
PID sensor and PCU-PCB is
well connected.
PID humid
PID humidity sensor error.
Relays
Relay does not work.
Restart the device.
Pump Motor
The pump motor does not work.
Check that the Pump motor
connector is well connected.
Pump Blocked
The pump was blocked or got heavy load.
Check sample gas input and
dust filter.
Check the tubing.
Valve
Valve does not work.
Check that the valve
connector is well connected.
LED
LEDs does not work.
Restart the device.
Filter dirty
The filter cartridge is dirty.
Install a new filter cartridge.
If the error code is still shown after a restart and troubleshooting, the device could be defective.
For additional support, please contact the manufacturer.