Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 30 of 60
The
Manual Purge
mode can be selected at any time while the device is running.
How To
–
Setup Manual Purge
1.
Go to the Instrument Menu and select Purge
2.
Select “run” to start configuration
3.
Select “MANUAL”
4.
Select PURGE TIME
and press “run” to start purging
Select PURGE TIME
•
Adjustable between 1 min and 5 hours
After purging the device independently returns again to the normal
operation.