Stationary Photoionization Detector SPID2
User Manual
SPID2 - User Manual (en) - 1.11.docx
Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de
Version 1.11
Page 21 of 60
6.1.2.
Calibration Test
The calibration test verifies both:
•
SPAN Gas calibration [ SAMPLE PORT ]
•
ZERO Gas calibration [ FILTER PORT
1
or SAMPLE PORT
2
]
Note 1: If Automatic Calibration Mode was used.
Note 2: If Manual Calibration Mode was used.
How To
–
Start Calibration Test
1.
Go to the Instrument Menu and select CALIBRATION
2.
Select “run” to go to calibration screen
3.
Select TEST and press
“accept”
4.
Select SAMPLE PORT or FILTER PORT and press
“run”
Calibration TEST from SAMPLE PORT
•
Apply SPAN Gas to the Sample Port
•
The current measurement value
will be displayed for 3 minutes
Calibration TEST from FILTER PORT
•
The device takes ambient air
through the Filter Port
•
The current measurement value
will be displayed for 2 minutes