ACE INSTRUMENTS WARNS: EVEN IF
YOU GET A RESULT WITH THE MEAN-
ING “SAFE RANGE”, YOU SHOULD NEV-
ER DRIVE AFTER HAVING CONSUMED
ALCOHOL OR ENGAGE ANY OTHER
ACTIVITIES THAT MIGHT ENDANGER
YOUR HEALTH AND SAFETY OR THE
HEALTH AND SAFETY OF OTHERS.
Warning notices
Technical data
FLO
The breath sample was not strong enough for measurement.
After warming automatically, please blow strong and steady during sampling.
BAT
The battery power is low. Please replace the 2 batteries inside the device with 2 new
ones (type AAA, 1.5V, LR03, Alkaline batteries). Do not use rechargeable batteries!
CAL
If you test over 1.000 tests, the ACE III Basic will display “CAL” sign for 3 seconds
when warming up for test. This “CAL” sign will be provided the appropriate time
for calibration. Please send the device to Alcohoot New York, For: ACE Instruments
Services.
Err6
Sensor module empty – please check sensor module.
Err9
System error – please contact ACE:
[email protected], www.ace-instruments.net
Statement of the BAC
0.000 – 0.400% BAC OR 0.000 – 4.000‰ (per mill) OR
0.000 – 2.000 mg/l (the default setting of your device is BAC)
Accuracy
1. +/- 0.005% BAC (Blood Alcohol Concentration)
with reference to 0.005% BAC
2. +/- 0.050‰ (per mill) with reference to 0.5‰
3. +/- 0.025 mg/l with reference to 0.25 mg/l
Warm-up time
5 seconds up to max. 2 minutes
Sensor
Fuel cell sensor
Response time
Within 10 seconds
Energy supply
2 pieces AAA batteries (1.5V, LR03, Alkaline)
Dimensions of the device
50 x 105 x 15 mm
Weight
85 g (including batteries)
Ambient conditions
Operation: 5 – 40 °C
Storage: 0 – 50 °C
Calibration
Every 6 – 12 months
RZ_160414_ACE BAs.indd 13
14.04.16 12:45