![Accutrol Apex Premier Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/accutrol/apex-premier/apex-premier_installation-operation-and-maintenance-manual_3845433039.webp)
T-APM - Apex Premier F
UME
H
OOD
M
ONITOR
T-APM 0133MAN00015 - Rev J ECN - JUL 2017
Page 39 of 61
APPENDIX F:
Configuration and Calibration
Screens
RELAY OUTPUT 2
PURGE
NORMALLY OPEN
CLOSE ON ALARM
NORMALLY CLOSED
OPEN ON ALARM
SCREEN 8D
CONFIGURATION MENU
SELECT OPTION
TO CONFIGURE
CHANGE
PASSWORD
SCREEN 1F
SCREEN 2E
HIGH
ALARM
EXIT
LOW
ALARM
SCREEN 3E
SELECT HIGH VELOCITY
ALARM SETPOINT
+
-
160
ACCEPT
DISABLE
FPM
SCREEN 8E
SELECT LOW VELOCITY
ALARM CAUTION BAND
+
-
60
ACCEPT
RESET
FPM
CHANGE PASSWORD
+
-
12345
ACCEPT
SCREEN 2F
CONFIGURATION MENU
SELECT OPTION
Calibration
Menu
SCREEN 1G
SCREEN 1H
CONFIGURATION MENU
SELECT OPTION
TO CONFIGURE
ALARMS
SCREEN 1E
CONFIGURATION MENU
SELECT OPTION
Restore All Factory
Defaults
RESTORE DEFAULTS
ARE YOU SURE?
ACCEPT CANCEL
SCREEN 2H
SCREEN 1J
WARNING TEMPERATURE
NOT WITHIN CAL. RANGE
CONTINUE CANCEL
CALIBRATION MENU
SELECT OPTION
Options displayed in
this scrolling box
SCREEN 1K
2 POINT CALIBRATION
OPEN SASH TO THE FULL
OPEN POSITION, WITH THE
STOPS DEFEATED
PRESS TO ACKNOWLEDGE
SCREEN 2K
2 POINT CALIBRATION
PERFORM TRAVERSE
ACROSS ENTIRE FACE AREA
OF FUME HOOD
PRESS TO ACKNOWLEDGE
SCREEN 3K
SCREEN 4K
2 POINT CALIBRATION
SELECT LOW FACE
VELOCITY
+
-
60
ACCEPT
RESET
FPM
CALIBRATING
PLEASE STAND BACK
WHILE DISPLAY
ACQUIRES DATA
IN 5 SECONDS
SCREEN 5K
SCREEN 1I
CONFIGURATION MENU
SELECT OPTION
EXIT
CANCEL
VERIFY NEW PASSWORD
+
-
12345
ACCEPT
SCREEN 3F
SCREEN 4E
SELECT HIGH VELOCITY
ALARM DELAY TIME
+
-
60
ACCEPT
RESET
SEC
SCREEN 5E
SELECT HIGH VELOCITY
ALARM CAUTION BAND
+
-
60
ACCEPT
RESET
FPM
SCREEN 6E
SELECT LOW VELOCITY
ALARM SETPOINT
+
-
70
ACCEPT
DISABLE
FPM
SCREEN 7E
SELECT LOW VELOCITY
ALARM DELAY TIME
+
-
60
ACCEPT
RESET
SEC
ACQUIRING DATA
102
SCREEN 6K
2 POINT CALIBRATION
SET SASH TO
MINIMUM POSITION
PRESS TO ACKNOWLEDGE
SCREEN 7K