background image

2.  Product Diagram

2

7

6.5  Maximum Power Limit

The myVolt Touch Power Supply has a maximum power 
output of 80 Watts.

Per Ohms Law:  

Power (P) = Current (I) x Voltage (V)

If the maximum power is reached during an electrophoresis 

run, the non-constant parameter will be limited to prevent 
exceeding the maximum power output

.

CAUTION:  

It is not recommended to run the power supply 

at or near the maximum wattage, as this will cause the power 
supply and gel tank to get hot.  Most electrophoresis buffer 
concentration will not have such low resistance that the 
current and voltage level will come close to the maximum 
power limit.  

Touch Screen

Control and Display

4 Pairs of 

Electrical Outputs

Folding 

Front Feet

Recesses for 

Stacking 2 Units

AC Power Inlet

and Fuse

6.4  Automatic Cross Over

During an electrophoresis run, the power supply will attempt to 
maintain the constant voltage or amperage setting, depending on 
which parameter is set to “CONSTANT”.

If, during a run, the non-constant parameter (limit) is met due to 
the resistance of the load, the power supply will automatically 
“cross over” so the non-constant parameter will now be the 
constant and limiting parameter.

FRONT

REAR

Rating, Serial #

Label

In constant amperage mode, the power supply will output the set 

mAmp current parameter, and the voltage will fluctuate relative 
to the resistance of the load (buffer concentration in the attached 
electrophoresis tanks) up to a maximum of the voltage setting 
that has been chosen.

Содержание myVolt Touch

Страница 1: ...300V Power Supply Operating Manual a division of This product is for research only ...

Страница 2: ...s that could damage plastic Use caution when cleaning near the housing vents Do not allow liquid to enter the power supply 9 Troubleshooting Problem Causes Measures No display Power not connected Fuse blown Check cord connection and power supply Unplug and re attach power cord Check and replace fuse at power inlet Voltage or Amperage setting not reached Maximum wattage exceeded Resistance of the l...

Страница 3: ... sets of outputs that accept standard electrical leads with banana plugs and should only be used with electrophoresis tanks that incorporate sufficient safety features to prevent electric shock 1 7 Overview of Electrical Principles Definitions Voltage The amount of potential energy between two points in a circuit Current The time rate of flow of an electric charge measured in amperes Resistance A ...

Страница 4: ...trol and Display 4 Pairs of Electrical Outputs Folding Front Feet Recesses for Stacking 2 Units AC Power Inlet and Fuse 6 4 Automatic Cross Over During an electrophoresis run the power supply will attempt to maintain the constant voltage or amperage setting depending on which parameter is set to CONSTANT If during a run the non constant parameter limit is met due to the resistance of the load the ...

Страница 5: ...on constant parameter set as a limit Press the operation icon CONSTANT to switch between CONSTANT CURRENT and CONSTANT VOLTAGE running modes In constant voltage mode the power supply will output the set voltage and the mAmp current value will fluctuate depending on the resistance of the load buffer concentration in the attached electrophoresis tanks up to a maximum of the mAmp current setting that...

Страница 6: ...ght strong light corrosive gas high magnetic fields heaters fires and other heat sources 4 5 6 Operating guide 6 1 Touch Screen Control Up Down Value Adjust Buttons Adjustable Parameters Operation Icon Buttons Status 4 Installation Place the power supply on a smooth level surface in an area where it will not contact liquids or chemicals The myVolt Power Supply features Dual Orientation technology ...