20
TROUBLESHOOTING
Problem
Possible causes
Possible soluTion
The unit cannot
power on
Is the battery exhausted?
Replace the battery.
Is the battery installed correctly?
Insert the battery observing polarity.
Stimulation weak
or cannot feel any
stimulation
Gel pads are dried out or dirty.
Replace with new gel pads.
Gel pads cannot stick to skin well.
Reconnect the pads.
Stimulation is
uncomfortable
Intensity is too high.
Decrease intensity.
Is the device being operated according to the manual?
Please check the manual before use.
The skin becomes
red and/or you feel
a stabbing pain
Use the gel pads on the same site every time.
Re-position the gel pads.
The gel pads are not sticking to the skin properly.
Ensure the gel pads are securely placed on
the skin.
The gel pads are dirty.
Replace with new gel pads.
The surface of the gel pads is scratched.
Replace with new gel pads.
If the unit does not operate after taking these measures, contact Carex Health Brands.