Guidance and manufacture’s declaration – electromagnetic immunity
The ACRL-9100 is intended for use in the electromagnetic environment specified below.
The user of the ACRL-9100 should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.6 GHz
3 V RMS outside the
ISM band, 6 V RMS
in the ISM and
amateur radio
bands
10 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the ACRL-9100, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance