Operation Manual for LCD Digital Readouts

 

Page 28 of 53

 

Statistics 

 

This Function can perform some basic statistical analysis 
without the use of SPC data collection on a PC or other 
device. The Statistics Function can provide the following information: 
1.  Measurement Count. 
2.  Minimum Measurement. 
3.  Maximum Measurement. 
4.  Average Measurement in two modes: 

a.  Standard average (Measurement Sum ÷ Count) 
b. 

Trimmed average ((Measurement Sum - Max - Min) ÷ (Count - 2)) 

 
Set Programming Parameter 

Pr24

 

 = 

4

 to enable this function.   

F3 & F4

 are used to execute this function. 

 
 

To begin the Statistics Function: 

 

1.  Clear the statistics variables by pressing the 

F4

 key.  

If the variables are already clear, the LCD will display 

CLr

.  

If there are previous samples in memory, the LCD will display 

Cn

 representing 

the measurement count for 1 second followed by a non-zero number.   
Press the 

F3

 

key to clear the memory. The LCD will display 

CLr

 for 1 second 

and then return to normal measurement mode  
 
2.  Take a measurement and press the 

F3

 

key. The LCD displays 

Add

  

x

 

momentarily to indicate that the measurement has been added and how many 
(

x

) have been accumulated so far.  Continue to take additional measurements 

and press the 

F3

 

key after each measurement. 

 
3.  Review the Data: 
Press the 

F4

 

key and the LCD will display 

Cn

 for 1 second followed by a non-

zero number. This number represents the measurement

 Count

 

Press the 

F4

 

key again and the LCD will display 

Lo

 for 1 second followed by 

the 

Minimum 

Measured value.  

Then you may: 
a. Press 

F3

 

to clear statistics variables and exit. 

b. Press 

F4

 

to continue without clearing the variables. 

c. Press 

SEND

 to transmit the Minimum Measurement to the output port.  

 

Press the 

F4

 

key again and the LCD will display 

HI

 for 1 second followed by 

the 

Maximum 

Measured value.  

Then you may: 
a. Press 

F3

 

to clear statistics variables and exit. 

b. Press 

F4

 

to continue without clearing the variables. 

c. Press 

SEND

 to transmit the Minimum Measurement to the output port.  

Содержание ProScale 950 Series

Страница 1: ...Basic 950 Series General Purpose General Purpose E In Panel LCD Readouts With Firmware Version 3 1 and higher ProScale LCD Readouts OPERATION...

Страница 2: ...RAMETERS HAVE BEEN PRE CONFIGURED AT THE FACTORY FOR THIS READOUT PARAMETER 2 Reading Direction __________________ PARAMETER 13 Linear Multiplier __________________ PARAMETER 14 ProScale Compatibility...

Страница 3: ...n this warranty Accurate Technology Inc does not make any warranties with respect to the product either expressed or implied including implied warranties of merchantability or fitness for a particular...

Страница 4: ...dout 13 12 24VDC OPERATION 14 General Purpose Readout 14 In Panel Readout 14 SECTION 3 OPERATION 15 INITIAL SET UP 15 READOUT KEY TIMING 16 READOUT KEY FUNCTIONS 17 On Off Key 17 Units Key 17 Plus Min...

Страница 5: ...PARAMETERS 39 PROGRAMMING PARAMETER SUMMARY 40 ENTER PROGRAMMING MODE 42 NAVIGATING PROGRAMMING MODE 42 EXIT PROGRAMMING MODE 43 PROGRAMMING PARAMETERS DETAILS 44 Pr1 Datum Key 44 Pr2 Direction 44 Pr3...

Страница 6: ...Linear Compensation 49 Pr40 Non Linear Compensation Interval 49 Pr41 Temperature Compensation 49 Pr42 Pr43 NO LONGER USED 49 Pr44 Temperature Compensation Coefficient 49 SECTION 5 FAQ 50 FREQUENTLY AS...

Страница 7: ...tance from some arbitrary point of origin DIGITAL READOUT or DRO is the component that displays the measurement or position information sent by the ENCODER The Readouts covered in this Manual are comp...

Страница 8: ...re version is displayed on power up i e P3 100 BASIC LCD Readout Part Number 700 1600 230 950 Series Readout Part Number 700 1600 M50 General Purpose GP LCD Readout Part Number 700 1600 235 General Pu...

Страница 9: ...Operating Power 950 Series Basic 1 CR123 3V Lithium Battery GP In Panel 1 CR123 3V Lithium Battery or 12 24VDC Operating Temperature 32 to 120 F 0 to 50 C Input 950 Series 10 pin ProScale ABS or Indu...

Страница 10: ...ting In Panel Mounting Software Battery Operation 12 24VDC Operation Programmable Datum Selectable Resolution Selectable Units Fractions Auto on off Offset Addition Upper Lower Limit Go NoGo Monitor P...

Страница 11: ...1600 235 SPC 701 1505 00x 700 1600 235 Programmable 701 1115 00x 700 1600 235 Panel Mount Battery 700 1600 435G 700 1600 435 Panel Mount Battery 700 1600 450 700 1600 435 Panel Mount Battery 700 1600...

Страница 12: ...side of the case Using any of the six holes on the back of the case M2 or 4 40 screws DIN In Panel A cutout should be made in the panel at least 90mm x 90mm 3 6 x 3 6 inches but no larger than 93mm x...

Страница 13: ...dout will turn itself off until the battery is replaced Basic 950 General Purpose Remove the screws in the upper right and lower left corners Pull the cover off Remove the old battery ies Install a ne...

Страница 14: ...out uses screw terminals for these connections Also See Section 3 Circuit Board Jumpers General Purpose Readout In Panel Readout When powering Readouts with 12 24VDC the Scale and the Readout common m...

Страница 15: ...t allows the current position to be easily configured using the and optionally DATUM Key Reference Point The ProScale Readout allows a reference point zero or other value to be easily recalled using t...

Страница 16: ...key has been pressed and held for the prescribed amount of time See the chart below for a summary of the functions performed by the keys below For a video demonstration of KEY TIMING click HERE Momen...

Страница 17: ...ight corner of the LCD Each bar represents 1 64th of an inch i e when in 1 16 fraction mode and three bars appear the measurement displayed is rounded down to the closest 1 16 inch plus each illuminat...

Страница 18: ...ogramming Parameter Pr3 Datum Key Momentarily pressing the DATUM key forces the Readout to a user programmed value This can be zero or any other displayable value Set the DATUM key programmed value wi...

Страница 19: ...ing Direction Resolution The Readouts have four selectable display resolutions Fractions display mode is not affected by resolution changes 1in 1mm 1cm or 01in 01mm 01cm or 001in 01mm 001cm or 0005in...

Страница 20: ...word LOCK will be displayed in the upper left corner of the LCD When LOCK is displayed on the LCD the DATUM and keys will not function to change the readout To de activate the Key Lock Press and Hold...

Страница 21: ...ause of one of these situations you can re adjust the Readout to correct the error This adjustment is referred to as the SEGMENT OFFSET ADJUSTMENT To add one segment to the displayed value Press and H...

Страница 22: ...mental when used in reference to the type of Measuring Technology employed by different ProScale products Absolute ABS The Readout automatically enters ABS mode when power is first applied This is ind...

Страница 23: ...a Wired Data Converter with Model 150 or 250 systems Scales and Encoders with part numbers ending in G should always be used Offset Addition Offset addition allows values to be pre programmed that are...

Страница 24: ...itor The Readout can monitor a position to detect drift or measurement variance To activate the Monitor mode position the Encoder system to the desired location and Momentarily press the F3 key The MO...

Страница 25: ...ature Momentarily press the F4 key again or cycle power NOTES If power is cycled when HOLD is active any key such as UNITS DATUM or that is depressed while in HOLD will be executed when power is resto...

Страница 26: ...points Pr25 while pressing F4 will decrement through the list When the point to be programmed is reached go to Step 2 2 Press and Hold the UNITS key and Momentarily press the F4 key The LCD will displ...

Страница 27: ...d i inches d If Pr25 is set to any non zero value and the Readout is in metric mode mm or cm the accumulated total is displayed as meters and millimeters MM mmm where M meters and m millimeters 2 Pres...

Страница 28: ...y The LCD will display CLr for 1 second and then return to normal measurement mode 2 Take a measurement and press the F3 key The LCD displays Add x momentarily to indicate that the measurement has bee...

Страница 29: ...ndard averaging will be influenced by the Minimum and Maximum measurement values If these values vary greatly from the measurement population the average will not reflect a centered representation of...

Страница 30: ...General Purpose Readout and the on board connector and circuit board pads on the In Panel Readout provide External Power connection points A solid state Output Signal connection 0 1A 30 VDC Auxiliary...

Страница 31: ...PROGRAMMING PARAMETER Pr37 The output is a transistor that conducts to Negative The factory default is normally open NO NOTES The LCD will flash when any of the programmed conditions above are encount...

Страница 32: ...circuit board There are three solder pads on both Readout circuit boards labeled IN1 IN2 and GND the In Panel readout connections are limited to circuit board solder pads no external connections Any...

Страница 33: ...ated for the different expansion rates based on a temperature sensor inside the Readout Use this feature to maintain the highest measurement accuracy possible when using the system under wide temperat...

Страница 34: ...ul Before beginning a Non Linear error compensation you will need to determine how many points and at what interval to compensate The Readout has a maximum of 126 correction points The smaller the com...

Страница 35: ...n the correction table calibration mode This alternating display will continue throughout the calibration process 6 Place the measuring system moving jaw stop against the fixed jaw stop Press DATUM to...

Страница 36: ...re process must be repeated Operating in Compensated mode After a successful correction table has been programmed the Readout will operate normally and the measuring system will have a higher level of...

Страница 37: ...e Digital Readout with specific parameters and preventing unauthorized or accidental changes JP4 Readout Power Selection Default Position B The General Purpose and In Panel Readouts operate on either...

Страница 38: ...Operation Manual for LCD Digital Readouts Page 38 of 53 General Purpose Basic 950 In Panel JP3 JP4 JP2 JP5 JP3 JP5 JP4...

Страница 39: ...release For Press Hold operations the Key Function is executed after the key has been depressed for the prescribed amount of time Resetting All Parameters To reset all programming parameters to their...

Страница 40: ...arameter Summary The table below lists all available Readout Programming Parameters and values that will be restored when the Readout is Reset To Factory Parameters Refer to Page 2 for any parameters...

Страница 41: ...0 to 16 0 Pr26 Drift Monitor Tolerance 01 mm to 3048 mm or 001 in to 120 in 010in Pr27 Upper Lower Limits 0 or 1 0 Pr28 Lower Limit 0 to 999 999in or 0 to 9999 99mm 0 Pr29 Upper Limit 0 to 999 999in o...

Страница 42: ...NITS key to advance up through the Programming Parameter list The Readout will display the Parameter number then the current programmed value To move down Parameter list Press and Hold the ON OFF key...

Страница 43: ...displays Pr x where x is the current parameter number 2 Momentarily press the DATUM key The LCD will display PG off 3 Release both keys The Readout will return to normal operation Video demonstration...

Страница 44: ...ection of Travel 0 or 1 This parameter controls the direction positive vs negative displayed on the readout when the measuring system is moved Default Value 0 Pr3 Key Lockout 0 or 1 This parameter con...

Страница 45: ...before the display turns off to conserve battery power The value is the number of minutes of idle operation no movement or key presses before the display turns off When this parameter is enabled 0 pr...

Страница 46: ...e it s operation See Advanced Programming Function for instances when this parameter is used Default Value 0 Pr26 Drift Monitor Tolerance 01 mm to 3048 mm or 001 in to 120 in This parameter sets the a...

Страница 47: ...m This value is added to the current measurement or position when Pr 30 0 and 1 is displayed in the upper left area of the LCD Default Value 500 in 12 7mm Pr32 Offset 2 0 to 999 999in or 0 to 9999 99m...

Страница 48: ...r 1 This sets the normal state of the programmable output when not activated and Circuit board jumper JP5 is in position A Default Value 0 Pr38 Programmable Output Function 0 to 3 Selects the function...

Страница 49: ...linear compensation calibration Default Value 5 000 in 127mm Pr41 Temperature Compensation 0 or 1 This parameter enables disables the temperature compensation feature Default Value 0 Pr42 Pr43 NO LON...

Страница 50: ...s message What does f err mean When the readout displays this message it means it is trying to display a fractional measurement out of the range typically over 400 inches Press the DATUM key twice the...

Страница 51: ...mitter at the Encoder sends data to a Receiver The receiver can be connected to a Readout at a remote location or to a PC RS232 or USB port Not compatible with 950 Series systems ProMUX ProMUX 3 is an...

Страница 52: ...ained single axis measuring table QC station Leg sets and casters make this product easy to move around the manufacturing floor Temperature compensation and non linear error correction are built into...

Страница 53: ...u for choosing a ProScale Product IT WAS PROUDLY MADE IN THE USA Accurate Technology Inc 270 Rutledge Rd Unit E Fletcher NC 28732 USA 828 654 7920 www proscale com Please register your ProScale produc...

Отзывы: