46
% saturation mode—
P7.0: Resetting to factory default
settings
This program lets you reset all parameters to factory default settings. This clears all
calibration data, memory, and any other setup functions you might have changed.
The LCD contrast setting will be reset to 8 (the maximum setting). However, the
following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• the temperature offset calibration value
• pH calibration data and set parameters
A
B
8.12
From measurement mode:
1.
Press the MODE key
to select the
% saturation measurement mode.
2.
Press SETUP key
to enter SETUP mode.
3.
Press the
▲
and
▼
keys
to scroll through
subgroups until you view parameter P7.0
in the lower display.
See figure
4.
Press the ENTER key
.
See figure
5.
Press the
▲
and
▼
keys
to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
6.
Press the ENTER key
to confirm
selection and to return to measurement
mode.
Notes
To reset all pH data, see page 38.
To reset all mg/l (ppm) data, see page 55.
See page 65 for a table of factory default
settings.