Dissolved Oxygen calibration in mg/l (ppm) mode
Calibrating the meter to 100 % in saturation mode will also calibrate the mg/l mode
at the value in mg/l corresponding to 100% saturation. This should produce
acceptable results in most applications.
This meter also lets you make a calibration adjustment in mg/l (ppm) mode
without affecting your % saturation calibration.
To select between mg/l and ppm units, see page 52.
17
To calibrate in mg/l (ppm) mode:
1.
Calibrate 100% saturation per section 4.5
on page 15.
2.
Rinse the probe well with deionized
rinse water or rinse solution.
Wipe the
outside of the probe carefully (do not
touch the membrane).
3.
Dip the probe into a sample of known
oxygen concentration.
Wait for the
reading to stabilize. If the Ready
indicator feature is enabled, it will appear
when the reading is stable (see page 43
for Ready indicator setup).
NOTE: The sample must continuously flow
past the membrane for accurate
readings. A stirrer can keep the
solution agitated.
4.
Press the MODE key
to select the
mg/l (ppm) mode.
See figure
5.
Press the CAL key.
The CAL indicator
will appear above the primary display.
The primary display shows the current
value of the measurement and the sec-
ondary display shows the temperature.
See figure
6.
Press the
▲
and
▼
keys
to adjust
the reading.
7.
Press the ENTER key.
The meter
automatically calibrates to the
value you entered and returns
to Measurement mode.
See figure
4.6