40
Temperature
sensor
Anode
Membrane
lock
Membrane
O-ring
Membrane
cap
Probe
guard
Cathode
Membrane replacement
Replacement of the membrane is required only
when you cannot calibrate the probe, or if the
membrane is damaged.
Typical membrane damages are punctures
or wrinkles caused during measurements or
cleaning.
To replace the probe membrane:
1.
Pull off the probe guard.
2.
Unscrew the membrane cap from the probe.
3.
Hold the probe under a water tap and brush
away the white oxide on the cylindrical anode
with a stiff plastic brush—do not use metal
cleaning material.
4.
It the cathode has any deposits, remove them
with a light scouring powder. Do not polish
the cathode.
NOTE:
If you have purchased a replacement
membrane module with preinstalled
membrane, skip to step 12.
CONTINUED ON NEXT PAGE
9.2
A