Temperature Calibration
The built-in temperature sensor included in the probe is factory calibrated.
Calibrate your sensor only if you suspect sensor drift that may have occurred
over a long period of time or if you have a replacement probe.
1.
Switch the meter on. Press MODE to
select mg/l (ppm) Measurement mode.
2.
Press the CAL/MEAS key to enter
mg/l (ppm) calibration mode.
The
CAL indicator will appear above the
primary display.
See figure
3.
While in mg/l (ppm) calibration mode,
press the MODE key to enter tempera-
ture calibration mode.
The primary
display shows the last set temperature
reading and the secondary display
shows you the factory default
temperature value.
See figure
4.
Compare the primary display reading
to a NIST-traceable thermometer or
another thermometer known to be
accurate.
5.
Press the
▲
or
▼
keys
to adjust
the primary display reading to agree
with your temperature standard.
See figure
6.
Press the ENTER key
to confirm
temperature calibration and return to
Measurement mode.
Notes
To exit from Temperature Calibration mode
without confirming calibration,
DO NOT
press ENTER in step 6. Press CAL/MEAS
instead.
Temperature calibration is restricted to
±5°C from the factory default value
displayed during calibration
(shown in the
secondary display).