II 

4 Technical Specifications 

1) 

Display Format:

 OLED Display; 

      SpO

2

 Measuring Range: 

0% - 100%; 

      Pulse Rate Measuring Range:

 30 bpm - 250 bpm; 

      Pulse Wave Display:

 columniation display and the waveform display. 

2) 

Power Requirements:

 2 ×1.5V AAA alkaline batteries (or using the rechargeable battery instead), 

adaptable range: 2.6V~3.6V. 

3) 

Power Consumption:

 Lower than 30mA. 

4) 

Resolution:

 1% for SpO

2

 and 1 bpm for Pulse Rate. 

5) 

Measurement  Accuracy: 

±2%  in  stage  of  70%-100%  SpO

2

,  and  meaningless  when  stage  being 

smaller          than 70%. ±2 bpm or±2% (select larger) for Pulse Rate.   

6) 

Measurement  Performance  in  Weak  Filling  Condition:

  SpO

2

  and  pulse  rate  can  be  shown 

correctly when pulse-filling ratio is 0.4%. SpO

2

 error is ±4%, pulse rate error is ±2 bpm or ±2% (select 

larger). 

7) 

Resistance  to  surrounding  light:

  The  deviation  between  the  value  measured  in  the  condition  of 

man-made light or indoor natural light and that of darkroom is less than ±1%. 

8)  It is equipped with a function switch. The Oximeter can be powered off in case there is no finger in 

the Oximeter for 5 seconds. 

9) 

Optical Sensor

 

      Red light (wavelength is 660nm, 6.65mW)   

      Infrared (wavelength is 880nm, 6.75mW) 

5 Accessories 

  One hanging rope; 

  Two batteries (optional); 

  One User Manual. 

6 Installation 

6.1 View of the Front Panel 

 

 

 

 

 

Figure 2 Front view 

6.2 Battery 

Step 1. Refer to Figure 3 and insert the two AAA size batteries in the right direction. 

Step 2. Replace the cover.   

Please take care when you insert the batteries as improper insertion may damage the device. 

 

 

 

 

 

 

 

Figure 3 Batteries installation 

6.3 Mounting the Hanging Rope   

Step 1. Put the end of the rope through the hole. 

Step 2. Put another end of the rope through the first one and then tighten it. 

 

Figure 4 Mounting the hanging rope 

7 Operating Guide 

1)  Insert the two batteries in the correct direction, and then replace the cover. 

2)  Open the clip as shown in Figure 5.   

 

 

 

 

 

Figure 5 Put finger in position 

3) 

Put  the  patient’s  finger  into  the  rubber  cushions  of  the  clip  (make  sure  the  finger  is  in  the  right 

position), and then clip the finger. 

4)  Press the button once on front panel. 

5)  Do not shake the finger and keep the patient at ease during the process. It is not recommended for 

the patient to be moving. 

6)  Get the information directly from screen display. 

7)  The button 

  has three functions. When the device is power off, pressing the button can turn it 

on; when the device is powered on, pressing the button shortly can change direction of the screen; 

when the device is powered on, long pressing the button can change brightness of the screen. 

  Fingernail and the screen should be on the same side.

 

8 Repairing and Maintenance 

  Please change the batteries when the low-voltage displayed on the screen. 

  Please clean the surface of the device before using. Wipe the device with medical alcohol first, and 

then let it dry in air or clean it with dry clean fabric.   

  Use the medical alcohol to disinfect the product after use prevents cross infection the next time it is 

used. 

  Please take out the batteries if the oximeter is not in use for a long period of time. 

  The best storage environment of the device is - 40ºC to 60ºC ambient temperature and no higher 

than 95% relative humidity.   

  Users are advised to calibrate the device  periodically (or according to the calibrating program  of 

hospital). It also can be performed by the state-appointed agent or just contact us for calibration. 

  High-pressure sterilization cannot be used on the device.   

  Do not immerse the device in liquid. 

  It is recommended that the device should be kept in a dry environment. Humidity may reduce 

the life span of the device, or even damage it. 

9 Troubleshooting 

Trouble 

Possible Reason 

Solution 

The SpO

2

 

and 

Pulse Rate cannot 
be displayed 
normally

 

1. The finger is not properly 
positioned. 
2. The patient’s SpO

2

 is too low to 

be detected.   

1. Place the finger properly and try 
again.   
2. Try again; Go to a hospital for a 
diagnosis if you are sure the device 
works correctly. 

The SpO

2

 

and 

Pulse Rate are not 
displayed stably

 

1. The  finger  is  not  placed  inside 
deep enough. 
2. The  finger  is  shaking  or  the 
patient is moving. 

1. Place the finger properly and try 
again.   
2. Keep the patient calm and at rest. 

The device cannot 
be turned on

 

1. The batteries are dead or 
almost dead.   
2. The batteries are not inserted 
properly.   
3. The device is defective. 

1. Change batteries. 
2. Reinstall batteries. 
3. Please contact the local service 
center. 

The display is off 
suddenly

 

1. The device will power off 
automatically when it gets no 
signal within 5 seconds. 
2. The batteries are almost dead. 

1. Normal. 
2. Change batteries. 

 

10 Key of Symbols 

Symbol 

Description 

 

Type BF 

 

Warning 

– See User Manual 

 

The pulse oxygen saturation (%) 

PRbpm

 

Pulse rate (bpm)

 

 

The  battery  voltage  is  insufficient  (change  the  battery  in  time  to  avoid 

inexact measurements) 

 

1. No finger inserted 

2. An indicator of signal inadequacy 

 

Battery positive electrode 

 

Battery cathode 

 

1.  Power switch 

2.  Change direction of the screen 

3.  Change brightness of the screen 

SN 

Serial number         

 

Alarm inhibit 

 

WEEE (2002/96/EC) 

IP22 

International Protection 

 

11 Function Specification 

Display Information 

Display Mode 

The Pulse Oxygen Saturation(SpO

2

OLED 

Pulse Rate(PR) 

OLED 

Pulse Intensity (bar-graph) 

OLED bar-graph display 

Pulse wave   

OLED 

SpO

2

 Parameter Specification 

Measuring range 

0%~100%, (the resolution is 1%). 

Accuracy 

70%~100%:±2%, Below 70% unspecified. 

Optical Sensor 

Red light (wavelength is 660nm) 

Infrared (wavelength is 880nm)   

Pulse Parameter Specification 

Measuring range 

30bpm~250bpm (the resolution is 1 bpm) 

Accuracy 

±2bpm or±2% select larger 

Pulse Intensity 

Range 

Continuous bar-graph display, the higher display indicates a stronger pulse. 

Battery Requirement 

1.5V (AAA size) alkaline batteries × 2 or rechargeable batteries                                                                                                                                                                                                                               

Battery Useful Life 

Two batteries can work continually for 20 hours 

Dimensions and Weight 

Dimensions 

57(L) × 31(W) × 32(H) mm 

Weight 

About 50g (with the batteries) 

 