AccuMed CMS-50D Скачать руководство пользователя страница 1

   

 

 

User Manual 

Instructions to User 

Thank you very much for purchasing our product! 

This  Manual  is  written  and  compiled  in  accordance  with  the

 

council  directive  MDD93/42/EEC  for 

medical  devices  and  harmonized  standards.  The  Manual  is  written  for  the  current  Pulse  Oximeter.  In 

case of modifications and software upgrades, the information contained in this document is subject to 

change without notice. 

The  Manual  describes,  in  accordance  with  the  Pulse  Oximeter's  features  and  requirements,  main 

structure,  functions,  specifications,  correct  methods  for  transportation,  installation,  usage,  operation, 

repair,  maintenance,  storage,  etc.  as  well  as  the  safety  procedures  to  protect  both  the  user  and 

equipment. Refer to the respective chapters for details. 

Please  read  the  Manual  very  carefully  before  using  this  equipment.  These  instructions  describe  the 

operating  procedures  to  be  followed  strictly;  failure  to  follow  these  instructions  can  cause  measuring 

abnormalities,  equipment  damage  and  personal  injury.  The  manufacturer  is  NOT  responsible  for  the 

safety,  reliability,  and  performance  issues  and  any  monitoring  abnormality,  personal  injury,  and 

equipment damage due to user's negligence of the operating instructions. The manufacturer's warranty 

service does not cover such faults.   

Due to the continual re-inovation, the specific products you received may not be totally in accordance 

with the description of this User Manual. We would sincerely apologize for this.   

This product is medical device, and can be used repeatedly. Its life span is 3 years.   

WARNING: 

 

An uncomfortable or painful feeling may develop if using the device ceaselessly, especially 

for the microcirculation barrier patients. It is recommended that the sensor should not be 

applied to the same finger for over 2 hours. 

 

For the individual patients, there should be a more prudent inspecting in the placing 

process. The device cannot be clipped on the edema and tender tissue. 

 

The light (the infrared is invisible) emitted from the device is harmful to the eyes, so the user 

and the maintenance man, cannot stare at the light. 

 

Testee cannot use enamel or other makeup. 

 

Testee’s fingernail cannot be too long. 

 

Please peruse the relative content about the clinical restrictions and warnings. 

 

This device is not intended for treatment. 

The User Manual is published by our company. All rights reserved. 

1 Safety 

1.1 Instructions for Safe Operations 

  Check the main unit and all accessories periodically to make sure that there is no visible damage 

that may affect patient’s safety and monitoring performance about the cables and transducers. It is 

recommended that the device should be inspected once a week at least. When there is obvious 

damage, stop using the monitor. 

  Necessary maintenance must  be performed  by  qualified service  engineers  ONLY.  Users  are  not 

permitted to maintain it by themselves. 

  The  oximeter  cannot  be  used  together  with  devices  not  specified  in  User's  Manual.  Only 

accessories that are recommended by the manufacture can be used with this device. 

  This product is calibrated before leaving the factory. 

1.2 Warnings 

  Explosive hazard

—DO NOT use the oximeter in environments with flammable gas such as some 

ignitable anesthetic agents. 

  DO NOT use the oximeter while the testee is undergoing a MRI and CT. 

  People who are allergic to rubber cannot use this device. 

  The disposal of scrap instruments, their accessories, and packing (including battery, plastic bags, 

foam, and paper boxes) should follow the local laws and regulations. 

  Please  check  the  packing  before  use  to  make  sure  the  device  and  accessories  are  totally  in 

accordance with the packing list, or else the device may have the possibility of working abnormally.   

  Please don't measure this device with function test paper for the device's related information. 

1.3 Attentions 

  Keep  the  oximeter  away  from  dust,  vibration,  corrosive  substances,  explosive  materials,  high 

temperatures, and moisture. 

  If the oximeter gets wet, please stop using it. 

  When it is carried from a cold environment to a warm or humid environment, please do not use it 

immediately.   

  DO NOT operate keys on front panel with sharp materials. 

  High  temperature  or  high  pressure  steam  disinfection  of  the  oximeter  is  not  permitted.  Refer  to 

User Manual in the relative chapter for instructions on cleaning and disinfection. 

  Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with 

medical alcohol on soft material. Do not spray any liquid on the device directly. 

  When cleaning the device with water, the temperature should be lower than 60ºC. 

  As to the fingers which are too thin or too cold, it would probably affect the normal measure of the 

patients' SpO

2

 and pulse rate. Please clip a thick finger such as thumb and middle finger deeply 

enough into the probe. 

  Do not use the device on infants or neonatal patients. 

  The  product  is  suitable for  children  above  four  years  old  and  adults  (weight  should  be  between 

15kg to 110kg).   

  The device may not work for all patients. If you are unable to achieve stable readings, discontinue 

use. 

  The  update  period  of  data  is  less  than  5  seconds,  which  is  changeable  according  to  different 

individual pulse rate. 

  The  waveform  is  normalized.  Please  read the measured  value  when  the  waveform  on screen  is 

equably and steady-going. This measured value is optimal value and the waveform at the moment 

is the standard one. 

  If some abnormal conditions appear on the screen during the test process, pull out the finger and 

reinsert to restore normal use. 

  The device has a normal life span of three years since the first electrified use.   

  The hanging rope attached the product is made from non- allergy material. If particular groups are 

sensitive to the hanging rope, stop using it. In addition, pay attention to the use of the hanging rope. 

Do not wear it around the neck to avoiding cause harm to the patient. 

  The instrument does not have a low-voltage alarm function. It only shows the low-voltage. Please 

change the battery when the battery energy is used up. 

  When the parameter is specific, the instrument does not have an alarm function. Do not use the 

device in situations where alarms are required.   

  Batteries must  be  removed  if  the device  is  going  to  be  stored  for more than  one month,  or  else 

batteries may leak. 

  A flexible circuit connects the two parts of the device. Do not twist or pull on the connection. 

2 Overview 

The  pulse  oxygen  saturation  is  the  percentage  of  HbO

2

  in the  total  Hb  in the  blood,  so-called  the  O

concentration  in  the  blood.  It  is  an  important  bio-parameter  for  the  respiration.  For  the  purpose  of 

measuring  the  SpO

2

  more  easily  and  accurately,  our  company  developed  the  Pulse  Oximeter.  At  the 

same time, the device can measure the pulse rate. 

The Pulse Oximeter features small size, low power consumption, convenient operation, and portability. 

It  is  only  necessary  for  patients  to  put  one  of  their  fingers  into  a  fingertip  photoelectric  sensor  for 

diagnosis, and a display screen will directly show the measured value of Hemoglobin Saturation.   

2.1 Classification:   

Class II b, (MDD93/42/EEC IX Rule 10)

 

2.2 Features 

  Operation of the product is simple and convenient. 

  The  product  is  small  in  size,  light  in  weight  (total  weight  is  about  50g  including  batteries)  and 

convenient to carry. 

  Power  consumption  of  the  product  is  low  and  the  two  originally  equipped  AAA  batteries  can  be 

operated continuously for 20 hours. 

  The product will automatically be powered off when no signal is  detected by the product within 5 

seconds. 

2.3 Major Applications and Scope of Application 

The Pulse Oximeter can be used to measure human Hemoglobin Saturation and pulse rate through the 

finger,  and  indicate  the  pulse  intensity  by  the  bar-display.  The  product  is  suitable  for  use  in  family, 

hospital (ordinary sickroom), Oxygen Bar, social medical organizations, and also the measure of oxygen 

saturation and pulse rate. 

 

The product is not suitable for use in continuous supervision for patients.

 

  The problem of overrating would emerge when the patient is suffering from toxicosis which 

is  caused  by  carbon  monoxide.  The  device  is  not  recommended  to  be  used  under  this 

circumstance.

 

2.4 Environment Requirements 

Storage Environment 

    a) Temperature: -40ºC~+60Cº 

    b) Relative humidity: 

≤95% 

    c) Atmospheric pressure: 500hPa~1060hPa 

Operating Environment 

    a) Temperature: 10ºC~40ºC 

    b) Relative Humidity: 

≤75% 

    c) Atmospheric pressure: 700hPa~1060hPa   

3 Principle and Caution 

3.1 Principle of Measurement 

Principle of the Oximeter is as follows: An experience formula of data processing is established taking 

use of Lambert Beer Law according to Spectrum Absorption Characteristics of  Reductive Hemoglobin 

(Hb) and Oxyhemoglobin (HbO

2

) in glow & near-infrared zones. Operation principle of the instrument is: 

Photoelectric  Oxyhemoglobin  Inspection  Technology  is  adopted  in  accordance  with  Capacity  Pulse 

Scanning & Recording Technology, so that two beams of different wavelength of lights can be focused 

onto the human nail tip through a perspective clamp finger-type sensor. Then a measured signal can be 

obtained  by  a  photosensitive  element,  information  acquired  through  this  will  be  shown  on  screen 

through processing in electronic circuits and a microprocessor. 

 

 

 

 

Figure 1 Operating principle 

3.2 Caution 

1.  The  finger  should  be  placed  properly  in  the  device  (see  the  attached  illustration  of  this 

manual ,Figure 5), or else it may cause inaccurate measurement. 

2.  The  SpO

2

 

sensor and photoelectric receiving tube should be arranged in a way with the subject’s 

arteriole in a position there between

 

3.  The SpO

2

 sensor should not be used at a location or limb tied with arterial canal or blood pressure 

cuff or receiving intravenous injection. 

4.  Make sure the optical path is free from any optical obstacles like rubberized fabric. 

5.  Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, 

infrared heater, direct sunlight, etc. 

6.  Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy. 

7.  Testee cannot use enamel or other makeup.

 

 

3.3 Clinical Restrictions 

1.  As the measurement is taken on the basis of arteriole pulse, substantial pulsating blood flow of the 

subject  is  required.  For  a  subject  with  weak  pulse  due  to  shock,  low  ambient/body  temperature, 

major bleeding, or use of vascular contracting drug, the SpO

2

 waveform (PLETH) will decrease. In 

this case, the measurement will be more sensitive to interference. 

2.  For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green, 

and  acid  indigo  blue),  carbon  monoxide  hemoglobin  (COHb),  methionine  (Me+Hb),  or  thiosalicylic 

hemoglobin,  and  some  with  icterus  problem,  the  SpO

2

  determination  by  this  monitor  may  be 

inaccurate. 

3.  The drugs like dopamine, procaine, prilocaine, lidocaine, and butacaine may also be a major factor 

blamed for serious error of SpO

2

 measurement.   

4.  As  the  SpO

2

  value  serves  as  a  reference  value  for  judgment  of  anemic  anoxia  and  toxic  anoxia, 

some patients with serious anemia may also report good SpO

2

 measurement. 