applicable  RF  compliance  level  above,  the  MODEL  ABP801  should  be  observed  to  verify  normal  operation.  If 

abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the 

MODEL ABP801. 

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

 

Recommended separation distances between 

portable and mobile RF communications equipment and the MODEL ABP801

 

The MODEL ABP801 is intended for use in an electromagnetic environment in which radiated 

RF  disturbances  are  controlled.  The  customer  or  the  user  of  the  MODEL  ABP801  can  help 

prevent  electromagnetic  interference  by  maintaining  a  minimum  distance  between  portable 

and  mobile  RF  communications  equipment  (transmitters)  and  the  MODEL  ABP801  as 

recommended  below,  according  to  the  maximum  output  power  of  the  communications 

equipment. 

Rated 

maximum 

output power 

of transmitter 

W

 

Separation distance according to frequency of transmitter m

 

150 kHz to 80 MHz

 

80 MHz to 800 MHz

 

800 MHz to 2.5 GHz

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended 

separation  distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the 

frequency of the transmitter, where P is the maximum output power rating of the transmitter in 

watts (W) according to the transmitter manufacturer. 

NOTE  1  At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency  range 

applies. 

NOTE  2  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is 

affected by absorption and reflection from structures, objects and people. 

 

Содержание ABP801

Страница 1: ...WRIST BLOOD PRESSURE MONITOR Instruction Manual MODEL ABP801 www AccuMed com ...

Страница 2: ... OF THE PRODUCT 4 PARTS IDENTIFICATION 5 INSERT OR REPLACE BATTERIES 5 TIME AND VOICE ON OFF OF SYSTEM SETUP 6 UNIT CONVERSION mmHg kPa DISPLAY 6 WHO BLOOD PRESSURE CLASSIFICATION DISPLAY 7 ATTACHING THE WRIST CUFF 7 HOW TO MEASURE BLOOD PRESSURE 7 CARE AND MAINTENANCE 8 SPECIFICATIONS 9 TROUBLESHOOTING 10 ...

Страница 3: ...ere are listed for your safe and correct use of the unit so as to prevent injuries or damages to the device The icons and meanings are as follow Examples of signs The icon indicates prohibitions what you should not do Matters involving actual prohibitions are indicated by text or pictures in or near The left icon refers to general prohibition The icon indicates something that is compulsory what mu...

Страница 4: ... Do not disassemble repair or remodel the main unit or the wrist cuff of the blood pressure monitor This will cause the unit to malfunction Requests from Manufacturer Make sure the connection tubing is not kinked before measuring to avoid any injury to patient For any patient do not measure more than 3 times continuously Wait for at least 5 minutes between any two measurements otherwise it will ca...

Страница 5: ...and fat intake Have regular physical checkups Maintain proper weight 3 Why measure blood pressure at home Blood pressure measured at a clinic or doctor s office may cause apprehension and produce an elevated reading 25 to 30 mmHg higher than that measured at home Home measurement reduces the effects of outside influences on blood pressure readings supplements the doctor s readings and provides a m...

Страница 6: ...asurements may be impaired if this device is used near televisions microwave ovens X ray mobile phone equipment or other devices with strong electrical fields To prevent such interference use the monitor at a sufficient distance away from such devices or turn them off 4 Before using you should wash your hands 5 Do not measure on the arm which is being used by monitoring ME Equipment Otherwise it c...

Страница 7: ...f not the device will not work When LOW BATTERY mark blinks on the display replace all batteries with new ones Do not mix old and new batteries It may shorten the battery life or cause the device to malfunction LOW BATTERY mark does not appear when the batteries run out Battery life varies with the ambient temperature and may be shorter at low temperatures The batteries may leak and cause a malfun...

Страница 8: ...te hour minute Voice on off until the setting is completed is On is the Off Non talking model does not have this function or you may choose method No 2 to set the voice Method No 2 for voice setting Please press and hold the MEM button when the device is in the shutdown state to enter the voice setting Please press the MEM button again to choose ON or OF Press the ON OFF button to quit UNIT CONVER...

Страница 9: ...at a table Rest your wrist on the table Relax for about 5 to 10 minutes before measurement Raise your hand so that the wrist cuff is at the same level as your heart Remain still and keep quiet during measurement Do not measure right after physical exercise or a bath Measure your blood pressure at about the same time every day HOW TO MEASURE BLOOD PRESSURE 1 Fasten the wrist cuff according to the i...

Страница 10: ...al blood pressure monitor in the best condition and protect the unit from damage follow the directions listed below Keep the monitor in the storage case when not in use Do not fold the arm cuff too tightly The fabric fastener could touch the inner surface of the arm cuff and damage it Clean the monitor and cuff with a soft dry cloth Do not use any abrasive or volatile cleaners CAUTION Do not subme...

Страница 11: ...ating condition 5ºC 40ºC 15 RH 93 RH Atmospheric pressure 70kPa 106kPa Storage condition 20 ºC 55 ºC 0 RH 93 RH Atmospheric pressure 50kPa 106kPa Dimensions Approx 72 W X67 H X33 D mm Weight Approx 130g excluding batteries Classification Type BF Wrist circumference 13 5 19 5 cm Specifications may be changed without notice in the event of improvements being made 1 Type of protection against electri...

Страница 12: ...t cuff was held lower than your heart Keep the correct position remain still and measure again The wrist cuff was not attached properly You moved your body or spoke while measuring The systolic pressure value or diastolic pressure value too low The wrist cuff was held higher than your heart You moved your body or spoke while measuring ACCOMPANYING DOCUMENTS A Instructions for use 1 MODEL ABP801 ne...

Страница 13: ...t is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for ...

Страница 14: ...mined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures obj...

Страница 15: ...ransmitters and the MODEL ABP801 as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum ...

Отзывы: