TROUBLESHOOTING
If you have trouble in using the unit please check the following points first.
ERROR DISPLAY
POSSIBLE CAUSE
HOW TO CORRECT
Nothing is displayed when
you
push
the
POWER
button or
battery icon
flashes.
No batteries installed
Insert batteries
Batteries are dead
Replace new batteries
The batteries are installed
correctly
Insert battery in the
correct polarities
C
an’t normally increase
pressure
Check your wrist cuff for any
air leakage
Replace wrist cuff with
new one
Inflate pressure is too high
Re-measure or send
back to dealer to
re-calibrate pressure
Shaking while measuring
Hand or body shaking while
measuring
Keeping static and
correct gesture and
measure again
Battery icon on
Batteries have low power
Replace batteries and
measure again
The systolic pressure value
or diastolic pressure value
too high
The wrist cuff was held lower
than your heart
Keep
the
correct
position, remain still,
and measure again
The wrist cuff was not
attached properly
You moved your body or
spoke while measuring
The systolic pressure value
or diastolic pressure value
too low
The wrist cuff was held higher
than your heart
You moved your body or
spoke while measuring
ACCOMPANYING DOCUMENTS
A.Instructions for use
1.MODEL ABP801 needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS;
2.Portable and mobile RF communications equipment can affect MODEL ABP801.
B
.
Technical description
1. Warning that the use of accessories, transducers, and cables other than those specified
with the exception of transducers and cables sold by the manufacturer of the MODEL ABP801
as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the MODEL ABP801.
2. Warning that the MODEL ABP801 should not be used adjacent to or stacked with other