AM1751 Secondary Reference PRT User’s Guide

                 

 

 

Rev. 201609V3 

7/10

4  Care and Handling Guidelines 

1.

 

DO NOT subject the PRT to any physical shocks and vibrations.  

a.

 

When not using the PRT, keep it in a place that’s not prone 
to drop, slam, bang, vibration or other strong physical 
contacts. Use a protective box or a carrying case whenever 
possible. 

b.

 

When shipping the PRT, use protective box and other 
protective packaging materials to minimize mechanical 
shocks as much as possible. 

c.

 

When using dry blocks, make sure the well diameter is 
appropriate to allow the PRT move up and down smoothly. 
 

2.

 

DO NOT subject the PRT to any contaminations. 

a.

 

Keep the PRT as clean as possible. Avoid contaminations 
as much as possible. 
 

3.

 

DO NOT over heat.  

a.

 

Do not use PRT above the manufacture specified 
temperature range. 

b.

 

Do not expose the PRT handle and lead wires to extreme 
temperatures. 

 

 
    
 
 
 
 
 
 
 

Содержание AM1751

Страница 1: ...Rev 201609V3 AM1751 Secondary Reference Platinum Resistance Thermometer User s Guide Please download the latest version at www accumac com...

Страница 2: ...librations 3 2 Specifications 4 2 1 Specifications 4 3 General Operations 5 3 1 Connecting to the readout device 5 3 2 Drive Current 5 3 3 Stability of Readings 5 3 4 Immersion Requirements 5 3 5 Ther...

Страница 3: ...live component 5 Follow all other safety guidelines listed in this user s guide Cautions Follow these guidelines to avoid possible damage to the instrument 1 Avoid mechanical shocks DO NOT drop or sla...

Страница 4: ...e temperature or change of temperature It has wide applications for dry wells or temperature baths 1 2 Main Features High accuracy 0 015 C at 0 01 C Temperature range 200 C to 670 C Durable and shock...

Страница 5: ...005 C Self heating 50 mW C Response Time 9 seconds for 63 response to step change in water moving at 3 feet per second Measurement Current 0 5 mA or 1 mA Sensor Length 32 mm Sensor Location 5 mm from...

Страница 6: ...ach end of the sensor are identified by red black and white blue insulation 3 2 Drive Current AccuMac recommends 1mA as drive current to ensure the best measurement 3 3 Stability of Readings To achiev...

Страница 7: ...has gone through an annealing process and stability test to minimize the thermal EMF which is caused by either impurities of sensing element or temperature differentials at lead wires connection poin...

Страница 8: ...ssible b When shipping the PRT use protective box and other protective packaging materials to minimize mechanical shocks as much as possible c When using dry blocks make sure the well diameter is appr...

Страница 9: ...service Be sure to have the model number and serial number of your PRT available 5 2 Problem Causes and Solutions a R0 Rtpw becomes higher significantly This is likely caused by mechanical shocks The...

Страница 10: ...ty period begins on the date of the shipment Parts product repairs and services are warranted for 90 days The warranty extends only to the original buyer or end user customer of an AccuMac authorized...

Отзывы: