AccuMac AM1640 Скачать руководство пользователя страница 5

AM1640 Precision Industrial PRT User’s Guide

                     

 

 

Rev. 201609V3 

4/10

2  Specifications 

2.1  Specifications 

Temperature Range 

 -200 °C to 420 °C 

Resistance at 0 °C 

Nominal 100 Ω 

Temperature Coefficient 

0.00385 Ω/ Ω/°C 

Accuracy  

±0.04 °C at -200 °C 
±0.035 °C at 0 °C 
±0.05 °C at 200 °C 
±0.09 °C at 420 °C 

Drift 

±0.04 °C at 0 °C after 100 hours at 420 °C 

Short Term Stability 

±0.01 °C 

Thermal Shock 

±0.007 °C after 10 times thermal cycles from minimum to 
maximum temperatures 

Hysteresis 

<=0.01 °C 

Self-heating 

50 mW/°C 

Response Time 

5 seconds for 63% response to step change in water moving 
at 3 feet per second 

Measurement Current 

0.5 mA or 1 mA 

Sensor Length 

32 mm 

Sensor Location 

5 mm from tip 

Insulation Resistance 

>1000 MΩ at room temperature 

Sheath Material 

Inconel

tm

 

Dimension 

0.25 inch X 12 inch (6.35 mm X 305 mm) 

External Leads 

Teflon

tm 

– insulated copper wire, 4 leads, 2.5 meters 

Handle Dimension 

15 mm (OD) X 65 mm (L) 

Handle Temperature Range 

-50 °C to 180 °C 

Optional Calibration 

NIST traceable calibration and data available per request: 
1640-12: Ordering # 5014 

 

             

 

 
 
 
 
 

Содержание AM1640

Страница 1: ...Rev 201609V3 AM1640 Precision Industrial Platinum Resistance Thermometer User s Guide Please download the latest version at www accumac com...

Страница 2: ...librations 3 2 Specifications 4 2 1 Specifications 4 3 General Operations 5 3 1 Connecting to the readout device 5 3 2 Drive Current 5 3 3 Stability of Readings 5 3 4 Immersion Requirements 5 3 5 Ther...

Страница 3: ...live component 5 Follow all other safety guidelines listed in this user s guide Cautions Follow these guidelines to avoid possible damage to the instrument 1 Avoid mechanical shocks DO NOT drop or sla...

Страница 4: ...re temperature or change of temperature It has wide applications for dry wells or temperature baths 1 2 Main Features High accuracy 0 035 C at 0 C Temperature range 200 C to 420 C Durable and shock re...

Страница 5: ...m to maximum temperatures Hysteresis 0 01 C Self heating 50 mW C Response Time 5 seconds for 63 response to step change in water moving at 3 feet per second Measurement Current 0 5 mA or 1 mA Sensor L...

Страница 6: ...ach end of the sensor are identified by red black and white blue insulation 3 2 Drive Current AccuMac recommends 1mA as drive current to ensure the best measurement 3 3 Stability of Readings To achiev...

Страница 7: ...T has gone through an annealing process and stability test to minimize the thermal EMF which is caused by either impurities of sensing element or temperature differentials at lead wires connection poi...

Страница 8: ...ssible b When shipping the PRT use protective box and other protective packaging materials to minimize mechanical shocks as much as possible c When using dry blocks make sure the well diameter is appr...

Страница 9: ...service Be sure to have the model number and serial number of your PRT available 5 2 Problem Causes and Solutions a R0 Rtpw becomes higher significantly This is likely caused by mechanical shocks The...

Страница 10: ...ranty terms in the separate agreement shall control Each product from AccuMac Corporation is warranted to be free from defects in material and workmanship under normal use and service The warranty per...

Отзывы: