Immediately touch the tip of the test strip to the drop of control solution. The 
control solution is pulled into the test strip through the strip tip.   

 

Note: 

If the control solution sample does not fill the check window, do not add a second 
drop. Discard the test strip and start over with a new test strip. 

 

7. 

Hold it in the drop until the meter beeps, and then you see the meter count 
down on the screen and followed with your control test result after a beep. 

     

 

Note: Control results are not included in the 7, 14 and 30 day average calculation. 

 

Understand Your Control Test Result 

Compare your control test result with the ranges printed on the test strip vial label 
or on the foil pouch. 

 

Notes:   

If your control test result is out of range: 

 

Check the expiration dates and discard dates of the test strip and control solution. 
Make sure that the test strip vial has not been opened for more than 6 months and 
the control solution bottle has not been opened for more than 6 months. Discard any 
expired test strips or control solution.   

 

Confirm the temperature in which you are testing is between 10  and 40°C 

Содержание VGM06

Страница 1: ......

Страница 2: ...2017 AccuBioTech Co Ltd Number Effective Date...

Страница 3: ...stem is based on measurement of electrical current caused by the reaction of the glucose with the reagents on the electrode of the test strip The blood sample is pulled into the tip of the test strip...

Страница 4: ...ng the Lancing Device 8 Getting a Blood Drop and Testing 10 Discard the Used Test Strip 12 Expected Diabetes Control Goal 12 Removing the Used Lancet 13 Testing with Control Solution 15 Using the Mete...

Страница 5: ...1 CHARPTER 1 UNDERSTANDING YOUR TESTING TOOLS Your Meter System Overview The Accu Tell Blood Glucose Meter and Test Strip...

Страница 6: ...en presenting year or time Date month Indicates average value Indicates low battery and that replacement is necessary Center area on the display that shows test results or error codes Indicates the sy...

Страница 7: ...ure is standard in your country This unit of measure cannot be adjusted Do not get water or other liquids inside the meter Keep the strip port area clean Keep your meter dry and avoid exposing it to e...

Страница 8: ...onment should be evaluated prior to operation of this device Important Safety Information Always keep the test strips in the original vial Tightly close the vial immediately after you have removed the...

Страница 9: ...er in the year position will now flash on the display Press to adjust it then press and hold until the meter beeps to set Then it will shift to next digit for setting Repeat the above action until the...

Страница 10: ...The hour will now flashing Press to adjust the current hour press and hold until the meter beeps to set Then it will shift to next digit for setting Repeat the above action until the year setting is c...

Страница 11: ...the test strip into the meter in the direction of the arrows Meter turns on after a beep 4 A symbol with a test strip with a flashing blood will appear letting you know the meter is ready to test Not...

Страница 12: ...and push it until the lancet comes to a complete stop in the lancing device 2 Hold the lancet firmly in the lancing device and twist the safety tab of the lancet until it loosens then pull the safety...

Страница 13: ...ce cover There are a total of 11 puncture depth settings To reduce the discomfort use the lowest setting that still produces an adequate drop of blood Note Greater pressure of the lancing device again...

Страница 14: ...barrel back to set the lancing device You may hear a click to indicate the lancing device is now loaded and ready for obtaining a drop of blood 3 Press the lancing device against the puncture site to...

Страница 15: ...blood drop out of the punctures It is important to motivate the animals by means of positive reinforcement like Perform the blood glucose testing under calm and relaxed conditions Feed a Goodie as a r...

Страница 16: ...w Meter turns off automatically after a beep Potential Biohazard Dispose of the used test strips as medical waste Expected Diabetes Control Goal Cat Dog Non diabetic pet glucose 47 151 mmg dL 2 6 8 4...

Страница 17: ...ng from the test strip vial or foil pouch Make sure that you followed the test procedure correctly Perform a control solution test See Performing a Control Test for instructions After checking all of...

Страница 18: ...issing or loose when you take the lancet out of the bag Do not use the lancet if the needle is bent Be cautious whenever the lancet needle is exposed In order to reduce the risk of infection from prio...

Страница 19: ...t match with how you feel About the Control Solutions Only use Accu Tell Control Solutions Low Normal or High to practice on the system The control solution results are not included in the average val...

Страница 20: ...vial or foil pouch Tightly close the vial cap immediately after you have removed the test strip Note Check the expiration and discard dates of the test strips Do not use the expiredtest strip 4 Insert...

Страница 21: ...fter a beep Note Control results are not included in the 7 14 and 30 day average calculation Understand Your Control Test Result Compare your control test result with the ranges printed on the test st...

Страница 22: ...l be not included in your blood glucose averages Using the Meter Memory Your meter automatically stores up to 300 results with the time and date Test results are stored from the newest to the oldest T...

Страница 23: ...30 day average Continue to press to review previous results in order Results will be shown starting with the most recent Each result will show the date and time the test was taken When END appears on...

Страница 24: ...saved in memory If you take out the battery when your meter is turned off date and time setting will be saved for one minute insert the new battery immediately you do not need to reset the meter 4 Rem...

Страница 25: ...irt blood or control solution into the meter through the strip or data port It is recommended that you store the meter in the carrying case after each use The Accu Tell Blood Glucose Meter is a precis...

Страница 26: ...Move to an area that is within the operating range for the meter Let the meter adjust to this temperature for 20 minutes before performing a test Potential software or hardware issue Take out battery...

Страница 27: ...attery Battery life 12 months or approximately 1 000 tests Glucose units of measure The meter is pre set to either millimoles per liter mmol L or milligrams per deciliter mg dL depending on the standa...

Страница 28: ...lse results Consult your veterinary doctor to find your pet s hematocrit level Abnormally high levels of vitamin C and other reducing substances will produce false high blood glucose measurements The...

Страница 29: ...Manufacturer 2O C 30O C Temperature limitations Contains sufficient for n tests Use by Lot Number Authorized Representative Sterilized using irradiation Control range REF Catalog number Model Model nu...

Страница 30: ...r than obvious abuse within the first five 5 years from purchase we will replace it with a new meter free of charge For your records also write the purchase date of your product here Date of purchase...

Страница 31: ...n If the meter does not work for any reason other than obvious abuse in the first 5 years after purchase we will replace it with a new or equivalent meter free of charge Date of purchase Meter serial...

Страница 32: ...What type of diabetes you have _________________________________________________________ Suggestion Comments...

Отзывы: