EXI-300 INVERTED MICROSCOPE SERIES   

 

 

 
 

ACCU-SCOPE

®

     

                   

 

9

 

OPERATION

Plug the 3-prong line cord into the microscope and then into a grounded 120V  or 220V A.C. electrical outlet.  
Usage of a surge suppressor outlet is highly recommended. Turn the illuminator switch 

 

to “―” (Fig. 6)  For 

longer bulb life always turn the illuminator variable intensity knob to the lowest illumination intensity setting 
possible before turning the power on or off. 

 
 

 

Adjusting the Illumination 

(Fig. 6) 

The light level may need adjustment depending upon the 
specimen density and objective magnification.  Adjust the 
light intensity for comfortable viewing by turning the light 
intensity control knob clockwise to increase brightness.  
Turn counterclockwise to decrease brightness. 

 

 

Adjusting Interpupillary Distance

 (Fig. 7 & 8) 

To adjust the interpupillary distance, hold the left and right 
eyetubes while observing a specimen. Rotate the eyetubes 
around the central axis until the fields of view of both eyetubes 
coincide completely. A complete circle should be seen in the 
viewing field when viewing the specimen slide. An improper 
adjustment will cause operator fatigue and will disrupt the 
objective parfocality.   
 
Where the “

” 

 on the eyepiece tube lines up, that is the 

number for your interpupillary distance.  The range is 48-75mm.  
Be sure to write down you interpupillary number for future 
operation. 

 

 
 

 

 

 

Adjusting the Focus 

(Fig. 8) 

To ensure that you obtain sharp images with both eyes, (since 
eyes vary, especially for those wearing glasses) any eyesight 
variation can be corrected in the following manner.  Set both 
diopter collars 

 

to “0”. Using your left eye only and the 10X 

objective, focus your specimen by adjusting the coarse 
adjustment knob.  When the image is in view, refine the image 
to its sharpest focus by turning the fine adjustment knob.  
Rotate the diopter collar to obtain the sharpest focus. To obtain 
the same sharp image using your right eye, do not touch the 
coarse or fine adjustments.  Instead, rotate the right diopter 
collar until the sharpest image appears.  Repeat several times 
to check.

   

 

IMPORTANT:

 do not counter rotate the focusing knobs as this will 

cause severe problems and damage to the focusing system.

 

 

 

 

Fig. 7 

 

 

 

Fig. 8 

 

 

Fig. 6 

Содержание EXI-300 Series

Страница 1: ...MANUAL EXI 300 INVERTED MICROSCOPE SERIES...

Страница 2: ...S 5 ASSEMBLY DIAGRAM 6 DETAILED ASSEMBLY 7 8 OPERATION ILLUMINATION 9 INTERPUPILLARY DISTANCE 9 FOCUSING 9 ADJUSTABLE TENSION CONTROL 10 STAGE PLATES OPTIONAL 10 SELECTING THE LIGHT PATH 11 APERTURE D...

Страница 3: ...jectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt should be removed...

Страница 4: ...eded Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form Carefully remove the microscope from the EPE foam container by its arm and ba...

Страница 5: ...control knobs Camera Photo Port Tension Adjustment Collar Frame Stand Objective Nosepiece Coarse Fine Focus Adjustment Knob Base Eyepieces Filter Holder Phase Contrast Slider Optional Condenser Stage...

Страница 6: ...indicate the order of assembly Use the 1 5mm and 2mm hex wrenches that are supplied with your microscope when required Be sure to keep these wrenches for changing out components or making adjustments...

Страница 7: ...ectives from low to high magnification Fig 2 NOTE Always rotate the nosepiece by using the knurled nosepiece ring Keep the covers on any unused nosepiece openings to prevent dust and dirt from getting...

Страница 8: ...put voltage indicated on the rear label of the microscope corresponds to your line voltage The use of a different input voltage than indicated will cause severe damage to your microscope Connecting th...

Страница 9: ...te circle should be seen in the viewing field when viewing the specimen slide An improper adjustment will cause operator fatigue and will disrupt the objective parfocality Where the on the eyepiece tu...

Страница 10: ...ustment ring clockwise to loosen or counterclockwise to tighten according to user preference Fig 9 Using the Stage Plates Optional Fig 10 11 NOTE for optimal viewing ensure the thickness marked on eac...

Страница 11: ...t photo port Turning it counter clockwise to the closed position sends 100 of the light to the binocular viewing head Upper Knob Turning it clockwise all the way to the PHO position will send 80 of th...

Страница 12: ...f the N A of the objective in use However when observing a culture specimen which is not dyed set the iris diaphragm lever to 75 of the N A of the objective in use Phase Contrast Observation The EXI 3...

Страница 13: ...er is pre aligned at our facilities If realignment is necessary follow these steps 1 Place a specimen on the stage and bring it into focus 2 Replace the eyepiece in the eyepiece tube with the centerin...

Страница 14: ...ge contrast If this occurs repeat steps 1 5 for readjustment The centering procedure may have to be repeated in order to get the best possible contrast if a specimen slide or the bottom surface of a c...

Страница 15: ...AL PROBLEM CAUSE SOLUTION The illumination is on but the field of view is dark The socket pin is not connected to the illumination column Connect it securely The bulb is burnt out Replace it with a ne...

Страница 16: ...the condenser does not match the phase annulus of the objective Adjust the light annulus so that it matches the phase annulus of the objectives The light annulus and the phase annulus are not centere...

Страница 17: ...289 2228 for an authorized service center near you The light intensity is not enough Not use an designated lamp use an designated lamp The brightness adjustment knob is used wrong Adjust the brightnes...

Страница 18: ...haser LED lamps are warranted for a period of two years from the date of original invoice to the original end user purchaser The mercury power supply is warranted for a period of one year from the dat...

Отзывы: