EXC-100 MICROSCOPE SERIES
73 Mall Drive, Commack, NY 11725 • 631
-864-
1000 •
www.accu-scope.com
13
ADJUSTMENT & OPERATION
(continued)
Using Mechanical Stage
Fig. 15
Place the Specimen Slide
1. Push the lever
①
of the specimen holder backwards.
2. Place the slide on the stage with the cover glass
②
facing upwards.
Gently release the lever
①
and allow the clip to slowly close onto the
slide and clamp the slide firmly(see Fig.15).
★
Do not let go of the lever as the clamp will close too fast. This may
result in slide breakage and flying glass pieces.
3. Rotate the X and Y-axis knob
③
of the stage, and move the specimen
to the center (aligned with the center of objective).
Adjusting the Condenser
– Installing a Filter
Fig. 16
1. Before installing the filter, rotate the condenser to the bottom and
then open the filter holder.
★
Place the filter with the rough side downward.