3000-LED MICROSCOPE SERIES 

ACCU-SCOPE

®

3

 

SAFETY NOTES 

1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from

dropping and being damaged.

2. Do not discard the molded Styrofoam container; the container should be retained should the

microscope ever require reshipment.

3. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.

Ensure the microscope is located on a smooth, level and firm surface.

4. If any specimen solutions or other liquids splash onto the stage, objective or any other component,

disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.

5. All electrical connectors (power cord) should be inserted into an electrical surge suppressor to

prevent damage due to voltage fluctuations.

6. For safety when replacing the LED bulb or fuse, be sure the main switch is off (“O”), remove the

power cord, and replace the LED bulb after the bulb and the lamp house has completely cooled.

7. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The

use of a different input voltage other than indicated will cause severe damage to the microscope.

CARE AND MAINTENANCE

1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.

2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces

should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap
solution. Do not use organic solvents for cleansing.

3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from

an air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a
lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a
circular motion. A small amount of absorbent cotton wound on the end of a tapered stick such as
cotton swabs or Q-tips, makes a useful tool for cleaning recessed optical surfaces. Avoid using an
excessive amount of solvents as this may cause problems with optical coatings or cemented optics
or the flowing solvent may pick up grease making cleaning more difficult.  Oil immersion objectives
should be cleaned immediately after use by removing the oil with lens tissue or a clean, soft cloth.

4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not

in use.

5. ACCU-SCOPE

®

 microscopes are precision instruments which require periodic preventative

maintenance to maintain proper performance and to compensate for normal wear. An annual
schedule of preventative maintenance by qualified personnel is highly recommended. Your
authorized ACCU-SCOPE

 ® 

distributor can arrange for this service.

Содержание 3000-LED-SPH

Страница 1: ...MANUAL 3000 LED MICROSCOPE SERIES 100 Lauman Lane Suite A Hicksville NY 11801 Tel 877 877 7274 Fax 516 801 2046 Email Info nyscopes com www microscopeinternational com LVWULEXWHG E...

Страница 2: ...4 5 ASSEMBLY DIAGRAM PROCEDURE 6 8 ADJUSMENT OPERATION ILLUMINATION 9 MIRROR 9 SPECIMEN SLIDE 9 FOCUSING 10 INTERPUPILLARY DISTANCE 10 FIELD DIAPHRAGM 11 APERTURE DIAPHRAGM 11 OIL OBJECTIVE 12 FILTER...

Страница 3: ...ctives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt should be removed u...

Страница 4: ...t quality prior to shipment UNPACKING AND COMPONENTS Your microscope arrived packed in a molded Styrofoam container Do not discard the container the Styrofoam container should be retained for reshipme...

Страница 5: ...CROSCOPE SERIES ACCU SCOPE 5 COMPONENTS DIAGRAM Lock Screw Y Axis Adjustment Knob Coarse Focusing Knob Fine Focusing Knob X axis Adjustment Knob Power Switch Iris Diaphragm Adjustment Control Objectiv...

Страница 6: ...preassembled by our factory technicians at our New York facility prior to shipment Should you need to disassemble assemble your microscope in the future please follow the instructions outlined below...

Страница 7: ...oosen the condenser lock screw Insert the condenser into the hole of the stand according to the directional arrow until the condenser is even with the stand and then rotate the condenser to make sure...

Страница 8: ...r of power cord securely into the power socket see Fig 3 Insert the other connector securely into an electrical outlet The microscope uses a universal power supply so it may be used in any voltage ran...

Страница 9: ...power switch to the 0 off position Loosen the Allen screw with an Allen wrench to screw off the built in illuminator housing see Fig 5 Screw on the mirror according to the directional arrow and tight...

Страница 10: ...your left eye only and the 10X objective focus your specimen by adjusting the coarse adjustment knob When the image is in view refine the image to its sharpest focus by turning the fine adjustment kn...

Страница 11: ...r Illuminator Condenser Optional Adjusting the Aperture Diaphragm The aperture size is increased or decreased by rotating the condenser aperture diaphragm lever When the aperture is closed the brightn...

Страница 12: ...xR oil immersion objective is in the light path With your eye at the level of the stage use the coarse focus knob to raise the stage with the specimen cover glass When you see a flash of light at this...

Страница 13: ...ugh side downward onto the groove of the illuminator condenser Place the cover back on the illuminator condenser and re install the set screw Fig 13 Filters are available in clear blue frosted blue gr...

Страница 14: ...rn the main switch to 0 OFF before replacing the LED bulb Unplug the power cord from the microscope Gently place the microscope on its arm as shown in Figure 15 Using a Philips head screwdriver remove...

Страница 15: ...ulb in position and screw it to its mount Figure 19 Figure 18 above Reinstall the base plate and reconnect the make break connector with the connector from the new bulb Figure 20 Secure the base plate...

Страница 16: ...range 76x50mm Condenser Built in Koehler illuminator systems Pre centered Iris diaphragm Koehler illuminator condenser and mirror optional Illuminator Universal power supply 90 240V voltage 3W LED lig...

Страница 17: ...e objective correctly into the optical path Dirt or dust on the view field Dirt or dust on the lens eyepiece condenser objective collector lens or specimen Clean the lens Poor image quality No cover g...

Страница 18: ...dial and or iris diaphragm Insufficient brightness Lamp intensity is too low Aperture diaphragm closed too far Condenser position too low Adjust the light intensity by rotating the intensity control...

Страница 19: ...No electrical power Lamp bulb burnt out Fuse blown out Check power cord connection Replace bulb Replace fuse Slippage of focus when using the coarse focusing knob Tension adjustment is set too low Inc...

Страница 20: ...years from the date of invoice to the original end user purchaser The LED lamp is warranted for a period of two years from the date of invoice to the original end user purchaser This warranty does no...

Отзывы: