39
Meter Maintenance and Troubleshooting
7
(The error code and
Err
will alternate on the display.)
The meter’s internal clock may need to be reset.
Press the meter button. If the time and date were
set, you will need to reset them. See the
Time
and Date
section in the chapter
Your New
System
.
(The error code and
Err
will alternate on the display.)
The test strip may be damaged.
Repeat the blood glucose or control test with a
new test strip.
(The error code and
Err
will alternate on the display.)
Your blood sample may contain a high level of
ascorbate.
Contact your healthcare professional.
(The error code and
Err
will alternate on the display.)
The batteries are almost out of power.
Change the batteries now. If the message
reappears after the batteries have been replaced,
remove the batteries again, press and hold the
meter button for at least 2 seconds, then reinsert
the batteries.
54461_07947836001_07947836001_01_Inside.indb 39
10/7/16 10:50 AM
Содержание WX3-122
Страница 1: ...User s Manual Blood Glucose Meter 54461_07947836001_07947836001_01_COVER indb 1 10 3 16 8 02 AM ...
Страница 2: ...54461_07947836001_07947836001_01_COVER indb 2 10 3 16 8 02 AM ...
Страница 4: ...54461_07947836001_07947836001_01_Inside indb 2 10 7 16 10 50 AM ...
Страница 6: ...4 Introduction 54461_07947836001_07947836001_01_Inside indb 4 10 7 16 10 50 AM ...
Страница 12: ...10 Your New System 1 54461_07947836001_07947836001_01_Inside indb 10 10 7 16 10 50 AM ...
Страница 34: ...32 Connecting to a PC 6 54461_07947836001_07947836001_01_Inside indb 32 10 7 16 10 50 AM ...
Страница 48: ...46 Technical Information 8 54461_07947836001_07947836001_01_Inside indb 46 10 7 16 10 51 AM ...