99
Appendix
Fig. 5: Subsequent boluses
hyper
hypo
Target Value
Target Range
bG level
Bolus
Current
bG values
Where the measured value is greater than the currently allowed bG value, the
newly calculated bolus (light-blue) only considers the difference between the
current bG value (dotted line) and the currently allowed bG value (the height of
the blue area). During the Acting Time of the first correction bolus (the width
of the blue area), only the currently allowed bG value is used to calculate the
second bolus.