46
Maintenance and Troubleshooting
7
Display
Action
Not enough blood or control solution was drawn into the test strip for measurement or
was applied after the test had started. Discard the test strip and repeat the blood glucose
or control test.
Blood or control solution was applied to the test strip before the flashing drop symbol
appeared on the display. Discard the test strip and repeat the blood glucose or control
test.
An electronic error occurred, or in rare cases, a used test strip was removed and
reinserted. Turn the meter off and on, or take the battery out for 20 seconds and
reinsert it. Perform a blood glucose or control test.
The temperature is above or below the proper range for the system. Refer to the test strip
package insert for system operating conditions. Move to an area with the appropriate
conditions, wait 5 minutes, and repeat the blood glucose or control test. Do not artificially
heat or cool the meter.
The battery is almost out of power. Change the battery now. If the message reappears
after the battery has been replaced, remove the battery again, press any meter button,
then reinsert the battery.
The time and date settings may be incorrect. Make sure the time and date are correct
and adjust, if necessary.
51801_07021933001.indb 46
5/7/14 8:53 AM
Содержание Aviva Plus
Страница 2: ...51801_07021933001_CVR indb 2 5 7 14 9 12 AM ...
Страница 4: ...51801_07021933001 indb 2 5 7 14 8 52 AM ...
Страница 8: ...Important Safety Information 6 51801_07021933001 indb 6 5 7 14 8 52 AM ...
Страница 14: ...12 Your New System 1 51801_07021933001 indb 12 5 7 14 8 52 AM ...
Страница 22: ...20 The ACCU CHEK Softclix Lancing Device 3 51801_07021933001 indb 20 5 7 14 8 52 AM ...
Страница 44: ...42 Meter and Lancing Device Cleaning and Disinfecting 6 51801_07021933001 indb 42 5 7 14 8 53 AM ...
Страница 60: ...51801_07021933001_ES indb 2 5 7 14 9 12 AM ...
Страница 64: ...Información importante de seguridad 6 51801_07021933001_ES indb 6 5 7 14 9 12 AM ...
Страница 70: ...12 Su nuevo sistema 1 51801_07021933001_ES indb 12 5 7 14 9 12 AM ...
Страница 78: ...20 El dispositivo de punción ACCU CHEK Softclix 3 51801_07021933001_ES indb 20 5 7 14 9 12 AM ...
Страница 100: ...42 Limpiar y desinfectar el medidor y el dispositivo de punción 6 51801_07021933001_ES indb 42 5 7 14 9 12 AM ...
Страница 115: ...51801_07021933001_CVR indb 3 5 7 14 9 12 AM ...