Technical data
Farbe: P 249 – Prüfmittelnummer: 203
199
Technical data
13
Weight
approx. 147 g with lancing device, batteries
and test strip drum
Display window
OLED (
O
rganic
L
ight
E
mitting
D
iode) display
Automatic power-off
after 60 seconds or 5 minutes depending on
operating status
Power supply
2 batteries (type AAA, LR 03, AM 4 or Micro)
or 2 NiMH rechargeable batteries (type AAA)
Battery life
approx. 1000 results or approx. 1 year (less
when the brightness of the display is set to
level 3 (high), or in Acoustic Mode owing to
the higher power consumption, or when
rechargeable batteries are used)
Interface
Infrared interface
Safety class
III
LED/IRED
Class 1
Electromagnetic
compatibility
This device meets the electromagnetic immu-
nity requirements according to EN ISO 15197
Annex A. The chosen basis for the immunity
tests (on electrostatic discharge) was basic
standard IEC 61000-4-2.
In addition it meets the electromagnetic
emissions requirements according to EN
61326. Its electromagnetic emission is thus
low. Interference from other electrically pow-
ered equipment is not to be anticipated.
Performance evaluation
The performance data for the Accu-Chek
Compact Plus system (Accu-Chek Compact
Plus meter together with Accu-Chek Compact
test strips) were determined using capillary
blood from patients with diabetes (method
comparison, accuracy), venous blood (repeat-
ability) and control solutions (reproducibility).
The system is calibrated with venous blood
containing various levels of glucose. The ref-
erence values are obtained using the hexoki-
nase method. The method comparison was
conducted by comparing the results from the
system with those of the hexokinase method
with deproteination using an automatic ana-
lyzer. The hexokinase method is traceable to
a NIST standard (traceability).
The Accu-Chek Compact Plus system complies with the requirements of
EN ISO 15197.