Accora Configura Rise and Recline Скачать руководство пользователя страница 2

2

©  Accora  Configura®  assessment  helpline:  01223  206110  |  [email protected]          

Introduction

This User Guide applies to the  
Configura

®

 Rise and Recline chair by 

Accora.

Please read carefully before setting  
up and using the chair and keep in  
a safe place for future reference.

These instructions include details of the 
performance of the chair. If the chair 
malfunctions, shows signs of damage, 
excessive wear, or otherwise feel 
unsuitable, please take it out of service 
and contact your supplier immediately.

The Configura

®

 Rise and Recline should 

be regularly checked to ensure the  
setup is correct and it is working as 
intended for the user.

General

The Configura

®

 Rise and Recline chair is a 

Medical Device Class I product. 
 
 

The batch number and date of manufacture 
of this product should be recorded.

Accora Ltd is committed to deliver safe 
products to its customers and has a policy  
of continuous improvement.

You may see the following terms on this 
document:

WARNING

 is a statement that alerts the  

user to the possibility of injury, death or  
other serious adverse reactions associated 
with the use or misuse of the product.

CAUTION

 is a statement that alerts the 

user to the possibility of a problem with the 
product associated with its use or misuse. 
Such problems include product malfunction, 
product failure, damage to the product or 
damage to other property.

They may be conditions under 
which a product should not be used 
(contraindications). These situations  

generally exist when the risks associated  
with the use of the product are greater than 
the benefits.

Conditions of Use

It is essential to consult Accora in advance 
if you wish to use the Configura

®

 Rise and 

Recline chair for any purpose outside the 
use detailed in this manual.

The Configura

®

 Rise and Recline chair may 

only be used in an indoor environment.

 

 Caution, read the instructions 
before use.

  

For indoor use only

WARNING.

 Before using the chair, you must 

read the instruction manual, particularly 
the safety warnings and use this product 
in accordance with them. Failure to follow 
the warnings and instructions may result in 
serious injury.

WARNING.

 The chair is intended for interior 

use only.

WARNING.

 The chair should not be placed in 

an uneven surface.

Intended Environment

•  Domestic use
•  Care homes or nursing homes

Essential Performance

•  To provide comfortable seating

•   To raise a user from a sitting to a standing 

position

•   To lower a user from a standing to a 

sitting position

•  To tilt-in-space a user

CLEANING

RECOMMENDATIONS 

** The mattress cover should not be opened or taken off during cleaning.
** Laundering the mattress cover will invalidate Manufacturers Warranty.
** Using bleach as a cleaning agent will invalidate Manufacturers Warranty.

Carelessness causes fire. This product contains foam which

passes the specified test. All upholstery is cigarette resistant.

Weight Capacity: 160kg (25 stone)

YEAR                       

        193cm x 87cm

        200cm x 90cm

200cm x 105cm

Charter House • Barrington Road

Orwell • Cambridge • SG8 5QP

T: 01223 206100 • F: 01223 206120

www.accora.uk.com

Multicare

Mattress

Wipe with mild detergent. Wash or dry-clean 

using detergent up to 75

o

C. For disinfection 

use sodium hypo-chlorite at no more than 

1000 parts per million of active chlorine. 

Do not use biological or phenolic-based 

cleaners. Wash and dry completely before 

storage, Some surface wrinkling may occur.

JAN  FEB 

MAR  APR  MAY  JUN  JUL 

AUG  SEP  OCT  NOV  DEC

MULTICARE 1000 Comfort
MULTICARE 2000 Soft
MULTICARE 3000 Visco
MULTICARE 4000 Primo

Содержание Configura Rise and Recline

Страница 1: ...IFU CF1 001EN R EV 05 A P R 2018 Configura Rise and Recline Instruction Manual IMPORTANT READ CAREFULLY RETAIN FOR FUTURE REFERENCE...

Страница 2: ...ser to the possibility of a problem with the product associated with its use or misuse Such problems include product malfunction product failure damage to the product or damage to other property They...

Страница 3: ...n be easily re issued The chair configuration is prescribed by a healthcare professional Any changes to the specification of the chair should only be made after consultation with a healthcare professi...

Страница 4: ...manual WARNING The safety features and instructions for operating this product must be strictly observed WARNING Do not modify this product without the authorisation of the manufacturer WARNING This...

Страница 5: ...18 Large Low Profile Dual 22 16 17 18 20 18 Extra Large Low Profile DDualal 24 16 17 18 20 18 Depth Height Width CHAIR DIMENSIONS INCHES ORDER CODE Requires 1 Height Adjustment Kit KITHA 0 CF1 100 Re...

Страница 6: ...n performance to the Visco cushion 160kg 25 stone weight capacity Not available on Extra Small Visco std Prevention management Prevention Heat sensitive memory foam moulds to the user providing maximu...

Страница 7: ...fabric on armrests for increased comfort and pressure reduction Vapour Permeable fabric on Backrest Visco and Vapour Permeable fabric on backrest to provide increased comfort and pressure reduction Va...

Страница 8: ...able support Adjustable Lateral Support System Flat foam backrest with adjustable Velcro lateral support system Excellent for providing more aggressive lateral support and mild positioning 1 4 Depth A...

Страница 9: ...gura accora uk com Using the handset Single Motor Tilt in space Button 1 back up footrest down chair rises Button 2 chair lowers footrest up back down and tilts back Always warn the user before changi...

Страница 10: ...onfigura assessment helpline 01223 206110 configura accora uk com Dual Motor Tilt in space Button 1 back up footrest down chair rises Button 2 chair lowers footrest up back down and tilts back 2 4 1 3...

Страница 11: ...accora uk com Button 3 backrest reclines Button 4 backrest up Important notes The chair can be disabled by disconnecting the handset on the handset lead Please see picture It is very important that c...

Страница 12: ...You will find a small lever attached to the bracket which should be pushed forwards to lock the back into position Unlocked position Locked position The chair should be positioned with enough room for...

Страница 13: ...ING Do not reduce in excess the amount of filling material of the backrest as this could cause serious damage to the skin condition of the user If you have any question regarding this issue do not hes...

Страница 14: ...king clips on the base of the chair and pull the back upwards When moving or handling the chair DO NOT use the backrest use the base only Using the backrest may cause damage to the internal framework...

Страница 15: ...se may cause the fabric to snag If this occurs do not pull the thread carefully trim it with scissors Do not allow animals onto this furniture as their claws may cause damage to the fabric Animals als...

Страница 16: ...eck the wall socket is in working order by trying another appliance 3 Check the connection from the chair to the transformer 4 Check the chair is positioned with sufficient space to recline and is fre...

Страница 17: ...tact Accora Customer Service on 01223 206100 WARNING Only competent persons should perform maintenance on the chair WARNING If the user or the carer is aware that any parts of the chair are corroded d...

Страница 18: ...m Guarantee The warranty period for the Configura Rise and Recline is 24 months covering parts and labour only We require a purchase order for any parts despatched warranty or otherwise Any issues reg...

Страница 19: ...19 Accora Configura assessment helpline 01223 206110 configura accora uk com...

Страница 20: ...23 206120 info accora uk com www accora uk com Ashworth House 245 St Neots Road Hardwick Cambridge CB23 7QL T 01954 214040 F 01954 214060 info accora uk com www accora uk com Registered to ISO 9001 20...