Cleaning 

 

 

An air cushion system can be a source of infection and 
we recommend that you clean it regularly. 
 
1.

 

Cleaning the pump: 
 

 

Do not immerse or soak the pump unit. 

 

Check for external damage before cleaning. 

 

Place the pump on a work surface and spray or 
wipe the outside of the case with a suitable 
cleaning solution. 

 

Wipe case with a clean cloth. Make sure all 
areas are clean 

 top and bottom, both sides. 

 

Spray cloth with cleaning solution and clean 
faceplate. Do not allow excess cleaning solution 
on faceplate or control panel. Allow surface to 
thoroughly dry after cleaning. 

 

Do not use a Hypo carbonate or phenol based 
cleaning solution as this may cause damage to 
the case. 

 

Plug pump into the mains and test to check 
normal operation. 

 
2.

 

Cleaning the cushion cells: 
 

 

Brush off or wipe down all surfaces of the cover 
with soap and water before wetting with any 
liquid disinfectant. 

 

Covers are immersed and soaked in disinfectant 
for the required incubation time. 

 

After pre-soaking, the cover is rinsed through a 
regular cycle with no soap, then laundered with 
mild detergent. Wash on coldest setting. 

 

Covers are aerated until they are fully dry. 

 

Repeat the process with the tubing set: spray, 
incubate, and then wipe clean. 

 

The pump bag should be turned inside out and 
completely wiped down using disinfectant 
solution. 

 

Allow it to thoroughly air dry. Once the inside is 
dry, turn it back: wipe down the outside of the 
bag with disinfectant. 

 

Drying of all the components must be done in a 
sunless area. 
 

3.

 

Disinfection: 
 

 

For extreme soiling control use a 5% Sodium 
Hypochlorite (bleach) solution. 

 

12.

 

Storage and Transport 

 

 

 
Lay the cushion out flat and then roll it up. Do not fold, 
crease or stack the cushion. 
 
Unplug the pump and store with the proper 
identification tag.  
 

13.

 

Product Disposal 

 
Disposal of this product should be in accordance with 
local regulations. 
 

14.

 

Guarantee 

 
The warranty period for the Configura CushionAir 
system is 24 months, covering parts and labour only. 
We require a purchase order for any parts 
despatched, warranty or otherwise. Any issues 
regarding warranty should be addressed in the first 
instance to your supplier. 

 

 

WARNING 

 

Cleaning must be carried out by a competent 
person. 
 
When cleaning the cushion, use protective 
clothing such as gloves and aprons. Infection 
could be transmitted from user to user if the 
chair is reused, or from user to carer or 
cleaner. 
 
Before cleaning the cushion unplug the pump 
mains cable/charging cable from the socket.  
 
Avoid spillage of liquids while cleaning the 
product. 

 

All functionality of the cushion must be 
checked by a competent person after 
cleaning. 
 
 

WARNING 

 

Do not store or use the cushion or pump in 
damp or wet areas. 
 
Do not position the cushion or pump near or 
against sources of direct heat or naked 
flames. 
 
The cushion must be protected when being 
transported to ensure the air cells are not 
damaged. 
 
The pump must not be dropped and must be 
transported in a protective box. 
 

Содержание Configura CushionAir

Страница 1: ...Configura CushionAir System Instruction Manual Fits Configura Comfort Lite 2 and Advance IFU CA1 001EN REV 05 JAN 2021...

Страница 2: ...ilable to users and operators throughout the product service life If you need further information please contact us See section 16 for region specific contact details General This accessory is classif...

Страница 3: ...o not place or store the cushion or pump where it could fall into water Do not place the cushion or pump in any liquid or water Do not touch the cushion or pump if it has fallen into water Unplug it i...

Страница 4: ...ubject to professional assessment by a clinician Ambulant semi ambulant and non ambulant patients If existing pressure damage does not improve or the patient s condition changes the overall therapy so...

Страница 5: ...SC1 000 CUSPM 0 CF1 000 Mains powered pump KITCA 0 CM1 100 CUSCP 0 CC1 001 Battery powered pump KITCA 0 SC1 000 CUSCPB 0 CC1 000 Battery for battery powered pump CUSCP 0 CC1 001 KITCA 0 SC1 000 CUSCPC...

Страница 6: ...ed in line with the assessment recommendation Do not allow the user to sit in the chair until the air cells have fully inflated If an existing cushion system is being replaced with the CushionAir syst...

Страница 7: ...m the overlay 5 Insert the CushionAir cells into the overlay Make sure that the cells of the CushionAir face upwards and the tube set is fed out of the back of the chair through the opening in the foa...

Страница 8: ...tube set must be routed alongside the backrest fitting bracket 1 above tube 2 and then secured to the chair sidepiece 3 with the Velcro hook and loop fixing Figure 7 8b Configura Advance Figure 7 8c...

Страница 9: ...essure Indication The current pressure is shown by 3 LEDs green HIGH PRESSURE MEDIUM PRESSURE or LOW PRESSURE Mute Indication The MUTE LED yellow is ON when the MUTE mode is enabled and is OFF when th...

Страница 10: ...Put the pump in the pouch make sure the control panel is facing the transparent side of the pouch 5 Operation and set up 5 1 Before operating the CushionAir Battery Powered Pump ensure you have read a...

Страница 11: ...the tubes fully inflated and inspect for air leakage Check if there is any air leakage from cells Ensure no leakage occurs 6 2 The pump doesn t work Check if the battery charge is too low Charge the b...

Страница 12: ...chair 3 2 Connect the pump to a mains power supply and turn the pump ON When the CushionAir cushion is being inflated for the first time the low pressure light will be lit 3 3 Select the cushion firmn...

Страница 13: ...battery charger cord for any damage abrasions or excessive wear Check the pipe set for any damage abrasions excessive wear or kinks Check both pipes in the tube set are correctly fitted to the pump a...

Страница 14: ...is dry turn it back wipe down the outside of the bag with disinfectant Drying of all the components must be done in a sunless area 3 Disinfection For extreme soiling control use a 5 Sodium Hypochlorit...

Страница 15: ...Serial number Manufacturer Indoor use only Do not dispose of in domestic refuse Caution risk of electrical shock Type BF applied part Class II double insulated Medical Device in accordance with EU Me...

Страница 16: ...16 IFU CA1 001EN REV 05 JAN 2021 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T 44 0 1223 206100 info accora care www accora care...

Отзывы: