CYCLE / WALK PROFESSIONAL USER MANUAL
COPYRIGHT © 2011 - 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Operational Sequence
Follow the sequence below. The number in parenthesis references the location of the button from the control section
on the previous page.
1.
Install battery in compartment located on the back of the unit (9)
2.
Connect lead wires to corresponding bar code under the display window (green lead wire in the left input (6),
blue lead in the right input (7).
3.
Wash the skin with water, (use soap if dirty)
4.
Connect black unit lead wires to white electrode wires.
5.
Press the power on/off button (1) to power on the device.
6.
At power on, the display (8) will default to the pre-set program selected for your condition.
a.
If a second program has been ordered by your physician, press Program Button B (3) to access the
additional program.
b.
To return to the first program, press Program Button A (2)
7.
Apply electrodes to treatment area paying attention to the lead wire polarity (red and black tip pins)
8.
Adjust stimulation intensity to a gentle muscle action causing a minor contraction of the muscle. This will
sequence back and forth between the channels. This may be increased over time as the patient adapts to the current
to provide a stronger contraction. Note: It is not necessary to cause contraction intensities that create passive
movement during cycling or walking. The stimulation is intended to provide sensory input and comfortable motor
activation to re-educate muscle movement or treat disuse atrophy of the muscles.
a.
Increase / decrease (4) Output A. Press (+) button to increase intensity in channel A. Press the (-) button
to decrease intensity in channel A.
b.
Increase / decrease (5) Output B. Press (+) button to increase intensity in channel B. Press the (-) button
to decrease intensity in channel B.
NOTE
:
If electrodes are worn or not connected to the device, the highest output you will be able to obtain is
5mA. If this occurs, check the electrodes and wires. You may need tore-wet or replace the electrodes if they are
too dry.
Shortly after reaching the desired intensity setting, the unit will lock the setting preventing an accidental
increase in the intensity. A lock icon will be indicated on the display. To unlock, you must first press the
(-) button to decrease the intensity, and then press the (+) button to increase the intensity to the new
desired level.
9.
The treatment timer will automatically end the treatment in the time prescribed. Treatment ends when bars show
up on the display (8)
10.
Turn the unit off by pressing the Power Button (1)
11.
Remove the stimulation electrodes. Repack the re-useable electrodes into their foil pouch electrodes and follow
the directions on the label for electrode care and storage.
NOTE
:
If you are using single use electrodes, discard after each use.
12.
Clean unit with ACP Germicidal Wipes.
13.
Store unit in carrying case
14.
Record treatment activity and settings on form provided at the end of this manual.