32
CONDUCTIVITY
MEASUREMENT — AC-PHM7
TEMPERATURE COMPENSATION FACTOR
The default setting of the temp. compensation factor is 2.0% / °C. You can adjust this factor
based on test solution and experimental data in parameter setting P4. Refer to the "Parameter
Settings" section for more information.
The following table lists some common examples of setting up the temp. compensation factor.
Temperature Compensation Factor
2.12% / °C
1.72% / °C
1.88% / °C
1.32% / °C
0.96% / °C
Solution
NaCl
5% NaOH
Dilute ammonia
10% Hydrochloric acid
5% Sulfuric acid