– 11 –

5. Press the transducer against the probe-disc, making sure that the transducer

sits flat against the surface. The display should show some thickness value, and
the Stability Indicator should have nearly all its bars illuminated.

6. While the transducer is firmly coupled to the probe-disc, press the PRB-0 key

on the keypad. The TI-25M-MMX will display “Prb0” while it is calculating its
zero point.

7. Remove the transducer from the probe-disc.

At this point, the TI-25M-MMX has successfully calculated it’s internal error 
factor, and will compensate for this value in any subsequent measurements. When 
performing a “probe-zero”, the TI-25M-MMX will always use the sound-velocity 
value of the built-in probe-disc, even if some other velocity value has been
entered for making actual measurements. Though the TI-25M-MMX will remem-
ber the last “probe-zero” performed, it is generally a good idea to perform a
“probe-zero” whenever the gauge is turned on, as well as any time a different
transducer is used. This will ensure that the instrument is always correctly zeroed.

4.3 Calibration

In order for the TI-25M-MMX to make accurate measurements, it must be set to
the correct sound-velocity for the material being measured. Different types of 
material have different inherent sound-velocities. For example, the velocity 
of sound through steel is about 0.233 inches-per-microsecond, versus that of 
aluminum, which is about 0.248 inches-per-microsecond. If the gauge is not
set to the correct sound-velocity, all of the measurements the gauge makes will
be erroneous by some fixed percentage. The one point calibration is the simplest
and most commonly used calibration procedure -optimizing linearity over large
ranges. The two point calibration allows for greater accuracy over small ranges 
by calculating the probe zero and velocity. The TI-25M-MMX provides three sim-
ple methods for setting the sound-velocity, described in the following pages.

Note:

Although the TI-25M-MMX has a thru-paint/coating feature, one and two

point calibrations must be performed on material with the paint or coating
removed. Failure to remove the paint or coating prior to calibration, will result in
a multi material velocity calculation that may be different from the actual material 
velocity intended to be measured.

Calibration to a known thickness

Note:

This procedure requires a sample piece of the specific material to be meas-

ured, the exact thickness of which is known, e.g. from having been measured by
some other means.

1. Make sure the TI-25M-MMX is on and switched to P-E (pulse-echo) mode.

Press the Dual-Multi key to toggle modes.

– 18 –

9 .0  RS2 3 2  S

ERI AL

P

ORT

The TI-25M-MMX is equipped with an RS232 serial port. Using the accessory
cable (part# N-306-0010), the TI-25M-MMX has the ability to connect to a com-
puter, or external storage device. The following section outlines the procedure for
connecting the TI-25M-MMX to a computer, and how to collect data using any
standard communications program:

Connecting To a Computer

1. Connect the accessory cable (part# N-306-0010) to the 2 pin jack located on

the bottom of the TI-25M-MMX, and the 9 pin connector to a serial port on
the computer.

2. Start the communications software that will be used to collect the

measurements (i.e. Microsoft Windows 3.1 -Terminal, or 95, 98,
Me, XP -HyperTerminal).

3. Setup the communications software using the following parameters:

Data Bits -8, Parity -None, Stop Bits -1, Baud Rate 1200.

4. Set the communications software COMM port to the port number that the

TI-25M-MMX is connected.

5. After taking a measurement, press the SEND key to send the measurement

to the computer. The measurement will be displayed on the computer screen.

NOTE:

Communications software packages generally have the ability to capture

the screen data to a common text file. This text file, containing the measurements,
can then be imported into any common spreadsheet program (i.e. Excel, Quattro
Pro, Lotus123) for further reporting requirements.

Содержание TI-25MX

Страница 1: ...234 0451 Web https www abqindustrial net E mail info abqindustrial net Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail i...

Страница 2: ...g abuse or misapplication has taken place as determined by ELECTROMATIC All returns for warranty or non warranty repairs and or replacement must be authorized by ELECTROMATIC in advance with all repac...

Страница 3: ...ece of known thickness Ideally this sample material should be a part of the same piece being measured or at least from the same lam ination batch By calibrating to each test piece individually the eff...

Страница 4: ...creases In most applications with surface temperatures less than about 200 F 100 C no special procedures must be observed At temperatures above this point the change in sound velocity of the material...

Страница 5: ...eezing Point 20 F Vapor Pressure N A Evaporation Rate N A Specific Gravity 1 0Z Solubility in Water complete pH 7 35 7 9 Acoustic Imp 1 726x106 Vapor Density N A Appearance and Odor water white opaque...

Страница 6: ...M or a velocity value IN ms or M s The UP arrow key has two functions When the TI 25M MMX is in calibration mode this key is used to increase numeric values on the display An auto repeat function is b...

Страница 7: ...lay numeric values as well as occasional simple words to indicate the status of various settings When the TI 25M MMX is displaying thickness measure ments the display will hold the last value measured...

Страница 8: ...ting the probe supplied with the TI 25M MMX the user should perform a Probe Zero as described in Section 4 2 The Probe Zero Test Plate is used for this task It is located on the top edge of the gauge...

Страница 9: ...e factory calibrated for steel with an acoustic velocity of 0 2330 IN s 5920 M s To measure the thickness of any other material the calibration will have to be changed by adjusting the acoustic veloci...

Страница 10: ...this happens the TI 25M MMX may perform a measurement through this couplant film resulting in a measurement that is larger or smaller than it should be This phenomenon is obvious when one thickness va...

Страница 11: ...se echo mode by pressing the Dual Multi key to toggle the modes Note The Probe Zero feature is not used in Echo Echo Thru Paint mode and has been disabled If the PRB 0 key is pressed while in this mod...

Страница 12: ...ting the sound velocity described in the following pages Note Although the TI 25M MMX has a thru paint coating feature one and two point calibrations must be performed on material with the paint or co...

Страница 13: ...symbol is flashing press the CAL key again so that the IN ms or M s symbols are flashing 3 Use the UP and DOWN arrow keys to adjust the displayed velocity up or down until it matches the sound veloci...

Страница 14: ...pled repeat step 4 6 Press the CAL key The IN or MM symbol should begin flashing 7 Use the UP and DOWN arrow keys to adjust the displayed thickness up or down until it matches the thickness of the sam...

Страница 15: ...moved from the sample after a measurement the last reading will be retained on the display b Occasionally a small film of couplant will be drawn out between the probe and the surface as the probe is r...

Отзывы: