14
Precipitor 32
®
Troubleshooting guide
Problems
It displays nothing on the screen.
Noises from the machine.
The action of the magnetic rod
frame doesn’t work normally.
It doesn’t move down or move up.
The function of heating is abnormal.
The temperature doesn’t increase.
Stain on the magnetic rod.
The magnetic rod comes off the
magnetic rod frame.
Suggestions
Did you turn on the power system already?
Did the plug connect with the power system?
Measure the voltage of outlet.
Check the strip to see if it has been pushed
into the bottom of strip rack.
Check the plate to see if it has been pushed
into the bottom of plate rack.
Contact service center: 886-2-87511888.
Please make sure the two screws are
removed
before turning on t
he power.
Press the button “Stop” to termi
nate the
program. Then, press the button “Stop”
again to let the robot move back to the
initial site.
Run DEMO program without 96-deep well
plate. Check the magnetic rod frame to see
if it still is abnormal.
Contact service center: 886-2-87511888.
Check the heating switch to see if it has
been turned on.
Contact service center: 886-2-87511888.
Be sure to put a clean strip on the strip rack
whenever you operate it. Avoid to be stained
by the remaining reagents in the 96-deep
well plate.
Use clean and moisten cotton cloth or wet
tissues to wipe it out carefully. Do not use
any solvent or cleanser to clean it.
Contact service center: 886-2-87511888.