6.  Setting up the DIP switches 

Open the lid of the control unit by lifting it from the side (picture 9). Inside, you will find 
eight DIP switches that are used to select the desired functions (picture 4). Remember to 
select the desired settings using the DIP switches before using the device. 

The control unit has the following factory settings. DIP 1, 2, 4, 5, 6, 7 and 8 OFF (down). 
DIP 3 ON (up). With these settings: 

  Movement notification delay is 13 seconds. 

  The SW1 switch acts as the On/Off switch which means that the device can be 

switched on or off by pressing the switch for three (3) seconds. 

  The notification sound is loudest. 

  The bed exit notification is disabled. 

 

 

Remove the batteries and disconnect the device from the power supply and 
remove nurse call cable (if used) from connector X2 before opening the lid. 

 

Remove the batteries and disconnect the power supply before setting up the DIP 
switches in order to implement the new settings.  

 

6.1  Adjusting the tonic-clonic seizure notification delay (switches 1 & 2) 

When the device detects faster movement (between 3

–20Hz), its internal timer switches 

on. The device gives a notification if the movement continues for longer than the preset 
time. The desired delay should be set up according to the table below. 

The factory setting is 13 seconds. If the user makes other habitual movements, such as 
rocking, scratching or restlessness, a longer delay time can be selected to avoid false 
notifications. 

Setting the fast movements duration after 
which the notification goes off 

Switch 1 

Switch 2 

10 seconds  

ON (up)  

ON (up)  

13 seconds (default)  

OFF (down)   OFF (down)  

16 seconds  

ON (up)  

OFF (down)  

20 seconds  

OFF (down)   ON (up)  

 

 

 

Содержание Emfit Epilepsy Alarm

Страница 1: ...onitor Art no 464000 464031 Note The device must not be used if an operational failure of the device might lead to a delay in getting the appropriate treatment or medication for the patient which coul...

Страница 2: ...2...

Страница 3: ...6 3 SW 1 switch function switch 6 12 6 4 Adjusting the notification sound volume switches 7 and 8 12 7 Inserting batteries and battery service life 13 8 5V AC External power supply optional 14 9 Conne...

Страница 4: ...al of the device after use 23 21 Declaration of conformity EU 23 22 Abilia limited warranty statement 23 22 1 Your guarantee 23 22 2 Terms 23 22 3 Exclusions and limitations 24 22 4 Your legal rights...

Страница 5: ...nications or local area networks If this device is used with a pressure care mattress filled using a compression pump the device may not function normally in some cases Equipment is not suitable for u...

Страница 6: ...se chapter Electromagnetic conditions Symbol of European Waste Electrical and Electronic Equipment Directive WEEE Directive 2002 96 EC on waste electrical and electronic equipment in these instruction...

Страница 7: ...certified in the external power supply Product is C TICK certified in the external power supply Product is China RoHS 30 certified in the external power supply Product is BSMI certified in the externa...

Страница 8: ...consists of a bed sensor located under the mattress and a control unit where the bed sensor is connected Control unit should be placed on a table clipped on bed or mounted on wall The control unit no...

Страница 9: ...operator is the caregiver of the user The operator needs to be adult who fully understands instructions and limitations of use mentioned in these instructions for use Otherwise no special skills or k...

Страница 10: ...details picture 2 In the event of a power failure two high quality 1 5V AA alkaline batteries can be used as an emergency power supply Do not use any rechargeable batteries or lithiumion batteries Re...

Страница 11: ...er supply and remove nurse call cable if used from connector X2 before opening the lid Remove the batteries and disconnect the power supply before setting up the DIP switches in order to implement the...

Страница 12: ...switched on or off by pressing the switch for three 3 seconds OFF down The SW1 switch does not act as the On Off switch The device is always on when it is connected to the power supply or when the bat...

Страница 13: ...is someone in bed 75 of that time there are two notifications per day and sound notification is on 30 seconds each time When the batteries are getting low the red LED will begin to flash slowly A beep...

Страница 14: ...s connected the batteries act as backup power supply in the event of a power failure All alkaline batteries self discharge and start to leak when empty contaminating the device Ensure that batteries a...

Страница 15: ...the bed sensor picture 13 the power supply delivered as an accessory picture 14 and any connector cables picture 15 according to the pictures 9 1 X2 AUX Connector Pin Order picture 16 From left to rig...

Страница 16: ...place the bed sensor under the mattress or mattress topper never just under the sheet The bed sensor must not come into direct contact with a person The bed sensor is designed to last for a minimum o...

Страница 17: ...press the switch for too long you might accidentally switch off the device The notification sound will also stop when the fast paced movements stop 12 3Bed exit alarm bypass switch The SW1 switch act...

Страница 18: ...f No one is on the bed 13 2Blue Device on standby Light is off Device is switched off Light flashes slowly Device is switched on Light flashes rapidly a few times Device triggers a notification 13 3Re...

Страница 19: ...re high priority signals that will overrule information signals if both signals activate simultaneously 14 1Alarm signals Two high and low BEEPS in a loop Fast movements alarm goes off Three high and...

Страница 20: ...bed should not last long enough for the notification to sound Person weight Switch position in parenthesis if false alarms occur 75 kg 1 0 50 75 kg 2 1 35 50 kg 3 2 or 1 25 35 kg 4 or 5 3 or 2 15 25 k...

Страница 21: ...of the bed sensor under the mattress The correct position is at the chest height of the user and across the bed 3 Power supply battery check if power supply is used Make sure that power supply and bed...

Страница 22: ...e device on only after the person has fallen asleep If the bed exit alarm is enabled DIP switch 3 in DOWN position and sensitivity is not set high enough alarm may trigger Either disable alarm DIP swi...

Страница 23: ...ntact your dealer distributor or manufacturer To avoid any unnecessary inconvenience on your part we recommend you read these instructions for use carefully before seeking guarantee service 22 1Your g...

Страница 24: ...beyond specifications or features described in the instructions for use or modifications to the product to conform it to national or local technical or safety standards in countries other than those...

Страница 25: ...xample some countries prohibit the exclusion or limitation of damages resulting from negligence gross negligence wilful misconduct deceit and similar acts Abilia s liability under this guarantee will...

Страница 26: ...eset switch 8 pcs DIP switches for settings volume delay one 10 position rotary switch for adjusting sensitivity Signal lights 3 LEDs green blue and red Delays Movement notification delay alternatives...

Страница 27: ...W 15 W Output voltage 5 VDC Output current 1 2 A 3 0 A Electrical safety class Class II 23 4Ambient conditions Operating temperature 10 40 C Storage and transport temperature 30 50 C Relative humidity...

Страница 28: ...pecified below The customer or the user of the equipment should assure that it is used in such an environment RF emissions CISPR 11 Group 1 The Emfit Tonic Clonic Seizure Monitor uses RF energy only f...

Страница 29: ...61000 4 5 1 kV line s to line s IEC 60601 1 2 test level Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Страница 30: ...each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may...

Страница 31: ...to frequency of transmitter 150 kHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 5 GHz d 1 2 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m...

Страница 32: ...32 26 Appendix Related pictures...

Страница 33: ...33...

Страница 34: ...34...

Страница 35: ...35...

Страница 36: ...36 464000 464031_Manual Emfit Tonic Clonic Seizure Monitor_EN_RevF1 docx Dec 2017...