4-4
Troubleshooting
377-96 Error
Messages
Signal too weak
Multiple
♦
Increase the CCD gain to
four:
a.
Open 377-96 Collection.
b.
In the Run window select the
Run module.
c.
Double-click the small
document icon next to the
Run Module pulldown menu.
d.
Change the CCD gain to
four.
♦
Resuspend samples in less
volume (concentrate).
Problem
Possible Cause
Solution
Message
Possible Cause
Solution
A Valid 96 Lane Firmware
Image is Required!
A non-96 collection software
has tried to establish
communications with a 377
instrument that has the 96-lane
option installed.
Install the 96-lane collection
software and firmware.
EP Voltage Deviation Exceeds
Tolerance
The EP voltage deviated
outside its tolerance range. The
instrument operation is paused.
Call service.
Warning: Plate Out. Thermistor
P43/J43 Open/Short Circuit
Warning: Plate In. Thermistor
P44/J44 Open/Short Circuit
Warning: Possible Heater
Thermistor Open/Short Circuit
Indicates one of the following:
♦
Possible open or short
circuit exists with the
thermistor/cable connected
to J43 or J44.
♦
Temperature of the plate in
an instrument with the
100k ohm thermistors is
21.9 °C or less.
One of the thermistors is not
functioning properly.
Schedule a service call, and
continue to operate the
instrument as usual.
This message may appear
when you launch data
collection software and start a
plate check, prerun, or run.