3-2
Software and Firmware
Software
GeneScan Analysis
The current GeneScan Analysis software is v. 3.0 with GSGelTracker.
Sequence Analysis
The current Sequence Analysis software is v. 3.2 with SAGelTracker.
Collection
Software
The current Collection software for 96 lanes is v. 2.5.
Preference File
Define default values in the new software following the procedures
described under “Setting Preferences” in the ABI P
RISM
377 DNA
Sequencer User’s Manual.
The default values you define for Sequencing or GeneScan
run
modules and sample sheets are maintained in the Preference file.
CCD Pixel Position
The instrument is shipped with the correct CCD pixel position value in
memory. When a run is started, the software checks for a value greater
than zero. If the value is lost from memory, an error message is
displayed at the beginning of the run as shown below.
The CCD pixel position value may become corrupted as the result of a
power surge or power failure. If this occurs, you must enter the correct
value before starting the run. For details, refer to the ABI P
RISM
377
DNA Sequencer User’s Manual.