2 GETTING STARTED
Charging Status Light
Once the instrument is connected to an external power source and turned on
the yellow charging status light located on the front of the instrument will be
illuminated, indicating the charging status of the internal battery.
The charging status light will only be illuminated when the m-PIMA™
Analyser is connected to an external power source and turned on. It may
take several minutes until the instrument displays the correct charging
status.
In order to ensure maximum battery life, charge the internal m-PIMA™
Analyser battery whenever possible.
To recharge an empty battery takes approx. 90 minutes.
When your m-PIMA™ Analyser is running on battery power, the battery
status is indicated by the battery symbol within the display status line (see
page 21).
Process Status Light
On the right side of the m-PIMA™ Analyser door you find the process status
light. It is only illuminated during an active analysis process.
No cartridge inserted/
m-PIMA™ Analyser off
Off
Analysis in progress
Solid
Analysis done
Flashing
m-PIMA™ Analyser not connected to external power
Off
m-PIMA™ Analyser
battery charging
Flashing
m-PIMA™ Analyser
battery fully charged
Solid