SECTION 6
SECTION 7
HOW TO PERFORM PATIENT TESTING (CONT.)
41
i-STAT
Alinity — Quick Reference Guide
Art: 731848-01 Rev. M Rev. Date: 25-Jul-2022
When the test is complete, the test results are displayed as in the
example
above.
RESULTS
®
An audible cue will be heard when results are ready. Touch Silence or remove
cartridge to silence the audio.
®
The result page shown here is the default. The system administrator must
specify the ranges used in your facility.
®
The blinking page button at the bottom of the screen appears when here is more
than one page of results. All action tabs are inactive until he second page of
results has been viewed.
®
Occasionally, numeric results will be replaced with the following symbols.
When displayed, a new test must be performed.
<> – Instrument cannot calculate the result.
*** – (Starouts) Instrument is unable to determine a result.
A sample may also yield results that are preceded by a greater than (>) or less
than (<) symbol. These results are outside the instrument’s measurement range.
In order to determine the exact numeric result, the sample must be analyzed by
a different method.
AnGap, mmol/L
42
Na, mmol/L
129
BUN, mg/dL
48
Glu, mg/dL
274
K, mmol/L
3.1
Crea, mg/dL
3.8
iCa, mmol/L
0.94
Cl, mmol/L
81
TCO2, mmol/L
9
i-STAT CHEM8+
Pt: 123456
16FEB2025 09:45