![Abbott i-STAT Alinity Скачать руководство пользователя страница 32](http://html2.mh-extra.com/html/abbott/i-stat-alinity/i-stat-alinity_quick-reference-manual_3351098032.webp)
SECTION 6
28
i-STAT
Alinity — Quick Reference Guide
Art: 731848-01 Rev. M Rev. Date: 25-Jul-2022
Starting from the
Home Screen touch More Options
then
Quality
Options.
Return
Enable/Disable
Wireless
Quality Op�ons
Review Results
View Ac�on Ranges
And Analyte Info
Transmit Unsent
Results
Instrument Status
More Op�ons
Home
More Op�ons
16FEB2025 09:45
Home
Instrument Op�ons
Administra�ve
Op�ons
Next, touch the
Perform Electronic Simulator Test button.
Return
Cal/Ver
View Disabled
Cartridges
Quality Control
Perform
Proficiency Test
Perform Electronic
Simulator Test
Update eVAS
Quality Op�ons
Home
More Op�ons Quality Op�ons
16FEB2025 09:45
By carefully observing the text and graphic instruction, the user will be
able to successfully complete an Electronic Simulator test. In the event
that the test does not pass, follow the prompts on the screen.
Use care when handling the Electronic Simulator.
Avoid touching the sensor area. Replace cap after use.
HOW TO PERFORM QUALITY TESTING (CONT.)
ELECTRONIC SIMULATOR (CONT.)