![Abbott FreeStyle Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/abbott/freestyle/freestyle_owners-booklet_450908029.webp)
25
(Example only)
6.
View result .
The result appears on the display when the test is complete .
7.
Compare the control solution result to the range printed on the test strip vial .
The result should fall within the range .
Note: If you see an “Er1, Er2, Er3, or Er4” on the display, it is an error code . Go to
Error Codes section .
(Example only)
8.
Press and hold
c
for 2 seconds until
appears on the display to mark
the result as a control solution result . To unmark a control solution test result,
press and hold
c
for 2 seconds . A value with the
indicates you have
marked this reading as a control solution test .
9.
Remove the test strip to turn meter off .
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com