Message
E-6
E-1
E-9
What It Means
Meter error
No coding required
or
Test strip may be damaged,
used, or the meter does not
recognise it
Meter error
What To Do
1. Check that you are using the correct
strip for this meter. (See test strip
instructions for use to verify your strip
is compatible with this meter.)
2. Repeat the test using a test strip for
use with your meter.
3. If the error reappears, contact
Customer Service.
1. Check that you are using the correct
test strip for this meter. (See test strip
instructions for use to verify your strip
is compatible with this meter.)
2. Repeat the test using a test strip for
use with your meter.
3. If the error reappears, contact
Customer Service.
1. Turn off the meter.
2. Repeat the test using a new test strip.
3. If the error reappears, contact
Customer Service.