1
EN
G
LIS
H
1
Performing a Control Solution Test
Insert the top of the test strip into your meter
until it stops. This will power on the meter.
The code number on the display must match the
code number on the side of your test strip vial.
(Refer to “Setting the Test Strip Code”)
Wait
for the meter to display the Blood Drop and
Test Strip Symbols.
Note
:
Be ready to adjust the code on the meter to match the code
number on the test strip vial.
Gently touch only the
Sample Area
of the Test
Strip Edge to the control solution.
Important:
Do not apply control solution to both Test Strip Edges.
When the test strip is full. A short line moves clockwise around the
display screen while the sample is processing. If the sound volume
is on, your meter will beep.
ART06690-Rev-B-EN.indd 13